Heart Rhythm. 2005 Aug;2(8):837-41. doi: 10.1016/j.hrthm.2005.03.011.
State-of-the art cardiac pacemakers are protected against radiofrequency signals. Although there have been earlier clinical and in vitro reports of cellular phone interference with implantable devices, only a few studies have been performed in recent years. The ringing phase of digital GSM or PCS cellular phones includes a brief period of peak radiated power.
This study tested the protection offered by electromagnetic filters of cardiac pacemakers against cellular phone ringing.
We performed 330 consecutive tests in 158 patients at the time of routine examination in our pacemaker follow-up clinic. The programmed parameters remained unchanged before testing. During electrocardiographic monitoring, 2 single-band digital cellular phones consecutively placed over the pacemaker pocket each received a call. The phone systems tested were 1) GSM at a maximal power output of 2 W, operating on a 900 MHz carrier frequency, and 2) PCS at a maximal output of 1 W, operating on a 1800 MHz carrier frequency.
Interference was noted in only 5 tests, due to interaction by the GSM system with 4 unprotected pacemaker models. The GSM test was negative in 12 other tests of identical pulse generator models. The overall incidence of interference was 1.5% of tests.
Interference by cellular phone ringing occurred only with unprotected pacemaker models. Standard programming of these unprotected models was associated with a low incidence of interference.
先进的心脏起搏器可防止受到射频信号干扰。尽管早期有临床及体外研究报告称手机会干扰植入式设备,但近年来仅有少数研究。数字GSM或PCS手机的振铃阶段包含一段短暂的辐射功率峰值期。
本研究测试心脏起搏器的电磁滤波器对手机振铃干扰的防护作用。
在我们起搏器随访门诊的常规检查时,对158例患者连续进行了330次测试。测试前程控参数保持不变。在心电图监测过程中,两部单波段数字手机依次放置在起搏器囊袋上方并各接听一个电话。所测试的手机系统为:1)最大功率输出为2W、工作在900MHz载波频率的GSM手机,以及2)最大功率输出为1W、工作在1800MHz载波频率的PCS手机。
仅在5次测试中发现干扰,原因是GSM系统与4种无防护功能的起搏器型号之间存在相互作用。在相同脉冲发生器型号的其他12次测试中,GSM测试结果为阴性。干扰的总体发生率为测试次数的1.5%。
仅无防护功能的起搏器型号会出现手机振铃干扰。这些无防护功能型号的标准程控设置干扰发生率较低。