Retinal nerve fibre layer thickness profile in normal eyes using third-generation optical coherence tomography.

作者信息

Skaf M, Bernardes A B, Cardillo J A, Costa R A, Melo L A S, Castro J C, Varma R

机构信息

Glaucoma Section, Hospital de Olhos de Araraquara, Araraquara-SP, Brazil.

出版信息

Eye (Lond). 2006 Apr;20(4):431-9. doi: 10.1038/sj.eye.6701896.

Abstract

AIMS

To establish four normal retinal nerve fibre layer (RNFL) thickness radial profiles based on third-generation optical coherence tomography (OCT) and to compare them with previously reported histologic measurements.

METHODS

A total of 20 normal eyes were studied. A circular scan was adjusted to the size of the optic disc and three scans were performed with this radius and every 200 microm thereafter, up to a distance of 1400 microm. Four different radial sections (superotemporal, superonasal, inferonasal, and inferotemporal) were studied to establish RNFL thickness OCT profiles. Additionally, two radial scans orientated at 45 and 135 degrees crossing the optic disc centre were performed in six of 20 eyes, and RNFL thickness was measured at disc margin.

RESULTS

Quadrant location and distance from disc margin interaction in RNFL thickness was statistically significant (P<0.001). The RNFL thickness decreased (P<0.001) as the distance from the disc margin increased for all sections. The measurements automatically generated by the OCT built-in software were thinner (P<0.001) than histologic ones close to the disc margin.

CONCLUSIONS

Four normal OCT RNFL profiles were established and compared with histological data obtained from the same area. RNFL measurements assessed by OCT 3 were significantly thinner close to the optic disc margin.

摘要

目的

基于第三代光学相干断层扫描(OCT)建立四个正常视网膜神经纤维层(RNFL)厚度的径向剖面图,并将其与先前报道的组织学测量结果进行比较。

方法

共研究了20只正常眼睛。将圆形扫描调整到视盘大小,以该半径进行三次扫描,此后每隔200微米进行一次扫描,直至距离达到1400微米。研究四个不同的径向截面(颞上、鼻上、鼻下和颞下)以建立RNFL厚度的OCT剖面图。此外,在20只眼睛中的6只眼睛中进行了两次与视盘中心相交的45度和135度方向的径向扫描,并在视盘边缘测量RNFL厚度。

结果

RNFL厚度在象限位置和距视盘边缘距离的相互作用具有统计学意义(P<0.001)。所有截面的RNFL厚度均随着距视盘边缘距离的增加而降低(P<0.001)。OCT内置软件自动生成的测量值在视盘边缘附近比组织学测量值更薄(P<0.001)。

结论

建立了四个正常的OCT RNFL剖面图,并与从同一区域获得的组织学数据进行了比较。通过OCT 3评估的RNFL测量值在视盘边缘附近明显更薄。