Zhang Yue, Lü Li-na, Xu Ke-xin
College of Precision Instrument & Opto-Electronics Engineering, Tianjin Univeristy, Tianjin 300072, China.
Guang Pu Xue Yu Guang Pu Fen Xi. 2005 Apr;25(4):512-5.
The article puts milk forward as simulate sample for blood glucose measurement. With the new simulate sample, the authors make deep researches on measuring method and wavelength choice in the field of noninvasive blood glucose measurement by near-infrared diffuse reflectance spectroscopy. Depending on the absorption coefficient of lactose, the authors choose the two wavebands from 1560 to 1750 nm and form 2090 to 2190 nm. The model was built by partial least-squares (PLS), several pretreatment methods (including multiplicative scatter correction, first derivative and vector normalization) and different spectral regions were compared. The best correlation values for lactose were 0.99 and the best RMSEP values for lactose were 0.045. The results provide us with valuable theory reference and practical experiences and lead us to next research on noninvasive blood glucose measurement.
该文章提出将牛奶作为血糖测量的模拟样本。利用这种新的模拟样本,作者们对近红外漫反射光谱法在无创血糖测量领域的测量方法和波长选择进行了深入研究。根据乳糖的吸收系数,作者选择了1560至1750纳米以及2090至2190纳米这两个波段。通过偏最小二乘法(PLS)建立模型,并比较了几种预处理方法(包括多元散射校正、一阶导数和矢量归一化)以及不同的光谱区域。乳糖的最佳相关值为0.99,乳糖的最佳均方根误差值为0.045。这些结果为我们提供了有价值的理论参考和实践经验,并引领我们开展下一步无创血糖测量的研究。