Flow-injection determination of streptomycin residues in milk using the luminol-periodate-Mn2+ chemiluminescence system.

作者信息

Wan Guo-Hui, Cui Hua, Zheng Hai-Song, Pang Yong-Qiang, Liu Li-Juan, Yu Xiao-Feng

机构信息

Department of Chemistry, University of Science and Technology of China, Hefei, Anhui 230026, People's Republic of China.

出版信息

Luminescence. 2006 Jan-Feb;21(1):36-42. doi: 10.1002/bio.880.

Abstract

A novel flow-injection method with chemiluminescence detection was developed for the determination of streptomycin residues in milk, based on the enhancement by streptomycin of the luminol-potassium periodate-Mn(2+) chemiluminescence system. The assay allowed analysis of streptomycin residues in whole milk samples ( approximately 3.5% fat) after sample work-up procedures. The limit of detection was 5.16 x 10(-9) mol/L for milk, and was far below the regulations of maximum residue limits (MRLs). The relative standard deviation (RSD) for 8.66 x 10(-8) mol/L streptomycin was 1.94% (n = 15) in milk. The results obtained for the assay of streptomycin residues in milk was comparable with those obtained by the official methods (ELISA) and demonstrated good accuracy and precision. The possible mechanism is briefly discussed.

摘要

基于链霉素对鲁米诺-高碘酸钾-Mn(2+)化学发光体系的增强作用,开发了一种用于测定牛奶中链霉素残留的新型流动注射化学发光检测方法。该检测方法允许在样品预处理后分析全脂牛奶样品(约含3.5%脂肪)中的链霉素残留。牛奶中链霉素残留的检测限为5.16×10(-9) mol/L,远低于最大残留限量(MRLs)规定。牛奶中8.66×10(-8) mol/L链霉素的相对标准偏差(RSD)为1.94%(n = 15)。牛奶中链霉素残留检测结果与官方方法(ELISA)所得结果相当,具有良好的准确性和精密度。文中简要讨论了可能的机理。