Pointer Jonathan S
Optometric Research, 4A Market Square, Higham Ferrers, Northants NN10 8BP, UK.
Ophthalmic Physiol Opt. 2005 Sep;25(5):446-51. doi: 10.1111/j.1475-1313.2005.00303.x.
To attempt to improve the reliability of the Maddox Wing (MW) test for the measurement of horizontal heterophoria at near.
Horizontal heterophoria measurements at 30 cm using the MW instrument were made on a group of 58 normally-sighted binocular young adult subjects. The procedures were repeated at each of five weekly visits. The standard tangent scale card provided by the instrument's manufacturer was used as the control condition vs four alternative cards designed to enhance accommodative stimulation and improve operational reliability.
A statistically-similar mean near phoria value was recorded using each of the five cards. However one of the alternative card designs produced a small but statistically significantly less variable result than the others: this design featured a much smaller typeface (4pt) than that used for the digits on the standard card, and the horizontal row of numbers was flanked by two periods of a low spatial frequency (< 2 c deg(-1)) square wave grating.
It is suggested that for improved reliability of near phoria measurement when using the MW instrument a more appropriately detailed design of tangent card would promote enhanced accommodation, an operational consideration likely to be especially relevant when examining children.
尝试提高马多克斯杆(MW)试验在测量近距水平隐斜时的可靠性。
使用MW仪器对58名视力正常的双眼年轻成年受试者在30厘米处进行水平隐斜测量。在每周一次的五次就诊中每次都重复该程序。仪器制造商提供的标准正切标尺卡用作对照条件,与四张旨在增强调节刺激并提高操作可靠性的替代卡进行比较。
使用五张卡中的每一张记录的平均近距隐斜值在统计学上相似。然而,其中一种替代卡设计产生的结果变化较小,但在统计学上显著小于其他设计:这种设计的字体(4磅)比标准卡上的数字字体小得多,并且水平数字行两侧是两个低空间频率(<2 c deg(-1))方波光栅周期。
建议为提高使用MW仪器测量近距隐斜时的可靠性,更详细合适的正切卡设计将促进增强调节,这一操作考虑在检查儿童时可能特别相关。