Enhanced corneal compensation for scanning laser polarimetry on eyes with atypical polarisation pattern.

作者信息

Tóth M, Holló G

机构信息

Department of Ophthalmology, Semmelweis University, Budapest, Hungary.

出版信息

Br J Ophthalmol. 2005 Sep;89(9):1139-42. doi: 10.1136/bjo.2005.070011.

Abstract

AIM

To investigate the potential advantage of an enhanced corneal compensation algorithm (ECC) compared with variable corneal compensation (VCC) in the analysis of scanning laser polarimetric (SLP) images with atypical retardation pattern (ARP).

METHODS

SLP-VCC images with ARP (typical scan score (TSS) <80) of one eye of each of 27 glaucoma patients and 19 healthy subjects were compared with the corresponding SLP-ECC images obtained at the same session.

RESULTS

ARP was present in 10.4% of the normals and 15.5% of the glaucoma patients imaged with SLP-VCC over 9 months. In both groups TSS was higher for ECC than for VCC (p<0.001). In glaucoma TSNIT, superior and inferior average thickness values were significantly lower, and TSNIT (measuring ellipse around the optic nerve head in the four (temporal, superior, nasal, inferior) quadrants) standard deviation was significantly higher with ECC than with VCC (p<0.001). In the normal group nerve fibre indicator (NFI) was lower with ECC than with VCC (p=0.007). TSNIT average was smaller and TSNIT standard deviation was higher with ECC (p<0.001). Superior and inferior average thickness did not differ between VCC and ECC in the normal group.

CONCLUSIONS

The new ECC software substantially improves polarimetric image analysis on eyes showing atypical polarisation pattern.

摘要

目的

研究在分析具有非典型延迟模式(ARP)的扫描激光偏振(SLP)图像时,增强型角膜补偿算法(ECC)相较于可变角膜补偿(VCC)的潜在优势。

方法

将27例青光眼患者和19例健康受试者每只眼睛的具有ARP(典型扫描评分(TSS)<80)的SLP-VCC图像与在同一次检查中获得的相应SLP-ECC图像进行比较。

结果

在9个月内用SLP-VCC成像的正常受试者中,10.4%存在ARP,青光眼患者中这一比例为15.5%。在两组中,ECC的TSS均高于VCC(p<0.001)。在青光眼患者中,ECC测量的TSNIT、上方和下方平均厚度值显著更低,而TSNIT(在四个象限(颞侧、上方、鼻侧、下方)围绕视神经乳头测量的椭圆)标准差显著高于VCC(p<0.001)。在正常组中,ECC的神经纤维指标(NFI)低于VCC(p=0.007)。ECC测量的TSNIT平均值更小,TSNIT标准差更高(p<0.001)。在正常组中,VCC和ECC测量的上方和下方平均厚度无差异。

结论

新的ECC软件显著改善了对显示非典型偏振模式眼睛的偏振图像分析。

相似文献

1
Enhanced corneal compensation for scanning laser polarimetry on eyes with atypical polarisation pattern.
Br J Ophthalmol. 2005 Sep;89(9):1139-42. doi: 10.1136/bjo.2005.070011.
2
Diagnostic accuracy of scanning laser polarimetry with enhanced versus variable corneal compensation.
Ophthalmology. 2007 Nov;114(11):1988-93. doi: 10.1016/j.ophtha.2007.01.022. Epub 2007 Apr 24.
4
Enhanced imaging algorithm for scanning laser polarimetry with variable corneal compensation.
Invest Ophthalmol Vis Sci. 2006 Sep;47(9):3870-7. doi: 10.1167/iovs.05-0067.
5
Scanning laser polarimetry with variable and enhanced corneal compensation in normal and glaucomatous eyes.
Am J Ophthalmol. 2007 Feb;143(2):272-9. doi: 10.1016/j.ajo.2006.09.049. Epub 2006 Oct 25.
9
Scanning laser polarimetry with variable corneal compensation in primary angle-closure glaucoma.
Ophthalmology. 2008 Aug;115(8):1334-9. doi: 10.1016/j.ophtha.2007.10.042. Epub 2008 Jan 16.
10
Scanning laser polarimetry with enhanced corneal compensation in patients with open-angle glaucoma.
J Glaucoma. 2008 Jan-Feb;17(1):24-9. doi: 10.1097/IJG.0b013e318133fb47.

引用本文的文献

1
Normalization of Retinal Birefringence Scanning Signals.
Sensors (Basel). 2024 Dec 31;25(1):165. doi: 10.3390/s25010165.
2
Assessment of the retinal nerve fiber layer in individuals with obstructive sleep apnea.
BMC Ophthalmol. 2016 Apr 18;16:40. doi: 10.1186/s12886-016-0216-2.
4
Scanning laser polarimetry in glaucoma.
Indian J Ophthalmol. 2014 Nov;62(11):1045-1055. doi: 10.4103/0301-4738.146707.
6
Modern technologies for retinal scanning and imaging: an introduction for the biomedical engineer.
Biomed Eng Online. 2014 Apr 29;13:52. doi: 10.1186/1475-925X-13-52.
9
Associations with retinal nerve fiber layer measures in the EPIC-Norfolk Eye Study.
Invest Ophthalmol Vis Sci. 2013 Jul 26;54(7):5028-34. doi: 10.1167/iovs.13-11971.
10
The glaucoma detection capability of spectral-domain OCT and GDx-VCC deviation maps in early glaucoma patients with localized visual field defects.
Graefes Arch Clin Exp Ophthalmol. 2013 Oct;251(10):2371-82. doi: 10.1007/s00417-013-2362-z. Epub 2013 Jul 3.

本文引用的文献

5
Sensitivity and specificity of new GDx parameters.
J Glaucoma. 2004 Feb;13(1):28-33. doi: 10.1097/00061198-200402000-00006.
7
Glaucoma detection using scanning laser polarimetry with variable corneal polarization compensation.
Arch Ophthalmol. 2003 Feb;121(2):218-24. doi: 10.1001/archopht.121.2.218.
8
Effect of peripapillary chorioretinal atrophy on GDx parametersin patients with degenerative myopia.
Clin Exp Ophthalmol. 2002 Dec;30(6):411-4. doi: 10.1046/j.1442-9071.2002.00570.x.
9
Correction for corneal polarization axis improves the discriminating power of scanning laser polarimetry.
Am J Ophthalmol. 2002 Jul;134(1):27-33. doi: 10.1016/s0002-9394(02)01512-x.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验