Tóth M, Holló G
Department of Ophthalmology, Semmelweis University, Budapest, Hungary.
Br J Ophthalmol. 2005 Sep;89(9):1139-42. doi: 10.1136/bjo.2005.070011.
To investigate the potential advantage of an enhanced corneal compensation algorithm (ECC) compared with variable corneal compensation (VCC) in the analysis of scanning laser polarimetric (SLP) images with atypical retardation pattern (ARP).
SLP-VCC images with ARP (typical scan score (TSS) <80) of one eye of each of 27 glaucoma patients and 19 healthy subjects were compared with the corresponding SLP-ECC images obtained at the same session.
ARP was present in 10.4% of the normals and 15.5% of the glaucoma patients imaged with SLP-VCC over 9 months. In both groups TSS was higher for ECC than for VCC (p<0.001). In glaucoma TSNIT, superior and inferior average thickness values were significantly lower, and TSNIT (measuring ellipse around the optic nerve head in the four (temporal, superior, nasal, inferior) quadrants) standard deviation was significantly higher with ECC than with VCC (p<0.001). In the normal group nerve fibre indicator (NFI) was lower with ECC than with VCC (p=0.007). TSNIT average was smaller and TSNIT standard deviation was higher with ECC (p<0.001). Superior and inferior average thickness did not differ between VCC and ECC in the normal group.
The new ECC software substantially improves polarimetric image analysis on eyes showing atypical polarisation pattern.
研究在分析具有非典型延迟模式(ARP)的扫描激光偏振(SLP)图像时,增强型角膜补偿算法(ECC)相较于可变角膜补偿(VCC)的潜在优势。
将27例青光眼患者和19例健康受试者每只眼睛的具有ARP(典型扫描评分(TSS)<80)的SLP-VCC图像与在同一次检查中获得的相应SLP-ECC图像进行比较。
在9个月内用SLP-VCC成像的正常受试者中,10.4%存在ARP,青光眼患者中这一比例为15.5%。在两组中,ECC的TSS均高于VCC(p<0.001)。在青光眼患者中,ECC测量的TSNIT、上方和下方平均厚度值显著更低,而TSNIT(在四个象限(颞侧、上方、鼻侧、下方)围绕视神经乳头测量的椭圆)标准差显著高于VCC(p<0.001)。在正常组中,ECC的神经纤维指标(NFI)低于VCC(p=0.007)。ECC测量的TSNIT平均值更小,TSNIT标准差更高(p<0.001)。在正常组中,VCC和ECC测量的上方和下方平均厚度无差异。
新的ECC软件显著改善了对显示非典型偏振模式眼睛的偏振图像分析。