Schueler Beth A, Kallmes David F, Cloft Harry J
Department of Radiology, Mayo Clinic, College of Medicine, Rochester, MN 55905, USA.
AJNR Am J Neuroradiol. 2005 Sep;26(8):1898-901.
As the use of 3D rotational angiography (3D RA) for the evaluation of cerebral vasculature becomes more widespread, it is important to evaluate this imaging method's effect on patient radiation dose. The purpose of the study is to measure 3D RA radiation dose as compared with biplanar digital subtraction angiography (DSA).
The distribution and peak skin dose were measured for 3D RA and biplanar DSA by using an anthropomorphic skull phantom. In addition, the cumulative incident dose, summed over all images in each acquisition, was determined. Measurements were acquired for our facility's standard 3D RA acquisition mode (25 degrees /s rotational speed; 162 total frames) and other available acquisition mode selections.
For 3D RA, the skin dose was found to be distributed across the back and sides of the skull with the peak skin dose located at the center of the back of the skull. The peak skin dose for the standard 3D RA acquisition mode was 15 mGy. For a biplanar DSA run, the peak skin dose was 58 mGy, also located at the back of the skull. The cumulative incident dose for the standard 3D RA acquisition mode was 33 mGy, compared with 53 mGy for biplanar DSA.
The patient radiation dose for 3D RA is significantly lower than for biplanar DSA, by nearly a factor of 4 in peak skin dose and 40% lower in cumulative incident dose.
随着三维旋转血管造影(3D RA)在评估脑血管系统中的应用日益广泛,评估这种成像方法对患者辐射剂量的影响至关重要。本研究的目的是测量3D RA的辐射剂量,并与双平面数字减影血管造影(DSA)进行比较。
使用仿真人头骨模型测量3D RA和双平面DSA的体表剂量分布及峰值剂量。此外,还确定了每次采集的所有图像的累积入射剂量。对本机构的标准3D RA采集模式(转速25度/秒;共162帧)及其他可用的采集模式选项进行了测量。
对于3D RA,发现体表剂量分布在颅骨的后部和两侧,峰值体表剂量位于颅骨后部中央。标准3D RA采集模式的峰值体表剂量为15 mGy。对于一次双平面DSA检查,峰值体表剂量为58 mGy,同样位于颅骨后部。标准3D RA采集模式的累积入射剂量为33 mGy,而双平面DSA为53 mGy。
3D RA的患者辐射剂量显著低于双平面DSA,峰值体表剂量低近4倍,累积入射剂量低40%。