Biomechanical testing of quadriceps tendon-patellar bone grafts: an alternative graft source for press-fit anterior cruciate ligament reconstruction?

作者信息

Dargel Jens, Schmidt-Wiethoff Rüdiger, Schneider Thomas, Brüggemann Gert-Peter, Koebke Jürgen

机构信息

Institute for Biomechanics und Orthopaedics, German Sport University Cologne, Cologne 50927, Germany.

出版信息

Arch Orthop Trauma Surg. 2006 May;126(4):265-70. doi: 10.1007/s00402-005-0048-7. Epub 2005 Sep 17.

Abstract

INTRODUCTION

Press-fit fixation of bone-patellar tendon-bone (BPTB) grafts in anterior cruciate ligament (ACL) reconstruction has been analyzed biomechanically in previous studies; however, the use of quadriceps tendon-patellar bone (QTPB) grafts has not been studied so far. It is hypothesized that QTPB grafts provide primary fixation strength comparable to BPTB grafts in press-fit ACL reconstruction with respect to bone plug length and loading angle.

MATERIALS AND METHODS

Fifty-two QTPB grafts were harvested from fresh human cadaver knees (mean age 73.3 years) with the length of the patellar bone plug being either 15 mm (Group I) or 25 mm (Group II). The grafts were anchored within fresh porcine femora (mean age 12 months) using a press-fit fixation technique. Forty-eight specimens were loaded to failure at 10 mm/s with varying loading angles of 0 degrees , 30 degrees , and 60 degrees until failure. A microradiographic pre-post-implantation analysis was conducted on four grafts.

RESULTS

The biomechanical testing showed a significant difference in the ultimate failure loads comparing Group I (mean 224+/-79.3 N) to Group II (mean 339+/-61.4 N), both showing mean ultimate failure loads to increase with rising loading angle. The predominant mode of failure was graft pullout at axial loading and tendon rupture at 60 degrees loading angle. The microradiographic analysis revealed an iatrogenic damage of the bone-tendon junction on the cancellous aspect of the bone plug in all trials, corresponding with the site of impactor placement during implantation.

CONCLUSION

QTPB grafts provide a loading capability comparable to BPTB grafts in press-fit ACL reconstruction. The broad and profound area of quadriceps tendon attachment to the patellar bone plug makes graft implantation demanding.

摘要

引言

在先前的研究中已对前交叉韧带(ACL)重建中骨-髌腱-骨(BPTB)移植物的压配固定进行了生物力学分析;然而,迄今为止尚未对股四头肌肌腱-髌骨骨(QTPB)移植物的使用进行研究。据推测,在压配式ACL重建中,就骨栓长度和加载角度而言,QTPB移植物提供的初始固定强度与BPTB移植物相当。

材料与方法

从新鲜人尸体膝关节(平均年龄73.3岁)获取52条QTPB移植物,髌骨骨栓长度为15 mm(I组)或25 mm(II组)。采用压配固定技术将移植物固定于新鲜猪股骨(平均年龄12个月)内。48个标本以10 mm/s的速度在0度、30度和60度不同加载角度下加载直至失效。对4条移植物进行植入前后的显微放射分析。

结果

生物力学测试显示,I组(平均224±79.3 N)与II组(平均339±61.4 N)的极限破坏载荷存在显著差异,两组的平均极限破坏载荷均随加载角度增加而增加。主要的失效模式为轴向加载时移植物拔出以及60度加载角度时肌腱断裂。显微放射分析显示,在所有试验中,骨栓松质骨侧的骨-肌腱交界处均存在医源性损伤,这与植入过程中冲击器放置的位置相对应。

结论

在压配式ACL重建中,QTPB移植物提供的加载能力与BPTB移植物相当。股四头肌肌腱附着于髌骨骨栓的区域广泛且深入,使得移植物植入具有挑战性。