永久性心脏起搏器植入手术对患者和心脏病专家的辐射剂量。

Radiation doses to patients and cardiologists from permanent cardiac pacemaker implantation procedures.

作者信息

Tsalafoutas Ioannis A, Spanodimos Stavros G, Maniatis Petros N, Fournarakis George M, Koulentianos Elias D, Tsigas Dimitrios L

机构信息

Medical Physics Unit, Konstantopoulio-Agia Olga Hospital, 158 Kolokotroni, 162-31 Byronas, Athens, Greece.

出版信息

Pacing Clin Electrophysiol. 2005 Sep;28(9):910-6. doi: 10.1111/j.1540-8159.2005.00194.x.

Abstract

BACKGROUND

The purpose of this study was to determine the patient and cardiologist doses during the implantation of permanent cardiac pacemakers under fluoroscopic control.

METHODS

For 55 procedures concerning three different types of pacemakers (DDD, VDD, and VVI), the dose-area product (DAP) meter readings and fluoroscopy times were recorded. From these data, the dose to the operating cardiologist was estimated.

RESULTS

The median values of DAP and fluoroscopy time for all the procedures monitored were 11.4 Gycm(2) and 6.6 minute, respectively. For the 22 DDD, 21 VDD, and 12 VVI pacemakers implanted, the respective DAP median values were 14.7, 9.9, and 7.3 Gycm(2) and the respective median fluoroscopy times were 8.4, 5, and 2.9 minutes. The median doses to the hands, chest, eyes, and legs of the cardiologist conducting the manipulations were estimated to 0.21, 0.06, 0.03, and 0.11 mGy, respectively, per procedure.

CONCLUSIONS

Compared to the existing literature, the median DAP value of this study is almost identical to the 11.2 Gycm(2) reported from a sample of 627 patients in 17 different x-ray rooms, whereas the fluoroscopy times are within the range of values reported by other authors. Concerning the cardiologist exposure, the estimated values indicate that the implantation of pacemakers is a procedure that does not involve a severe risk, especially if it is taken into account that lead aprons and collars are routinely used.

摘要

背景

本研究的目的是确定在荧光镜控制下植入永久性心脏起搏器过程中患者和心脏病专家所接受的辐射剂量。

方法

针对涉及三种不同类型起搏器(DDD、VDD和VVI)的55例手术,记录剂量面积乘积(DAP)计读数和透视时间。根据这些数据,估算手术心脏病专家所接受的辐射剂量。

结果

所有监测手术的DAP和透视时间的中位数分别为11.4 Gycm²和6.6分钟。对于植入的22台DDD、21台VDD和12台VVI起搏器,各自的DAP中位数分别为14.7、9.9和7.3 Gycm²,各自的透视时间中位数分别为8.4、5和2.9分钟。估计每例手术中,进行操作的心脏病专家手部、胸部、眼睛和腿部所接受的辐射剂量中位数分别为0.21、0.06、0.03和0.11 mGy。

结论

与现有文献相比,本研究的DAP中位数与在17个不同X光室对627例患者样本所报告的11.2 Gycm²几乎相同,而透视时间在其他作者所报告的值范围内。关于心脏病专家所受辐射,估计值表明起搏器植入手术并非高风险手术,尤其是考虑到手术中常规使用铅围裙和铅颈围。