Meeks Sanford L, Harmon Joseph F, Langen Katja M, Willoughby Twyla R, Wagner Thomas H, Kupelian Patrick A
Department of Radiation Oncology, M D Anderson Cancer Center Orlando, Orlando, Florida 32806, USA.
Med Phys. 2005 Aug;32(8):2673-81. doi: 10.1118/1.1990289.
Helical tomotherapy is an innovative means of delivering IGRT and IMRT using a device that combines features of a linear accelerator and a helical computed tomography (CT) scanner. The HI-ART II can generate CT images from the same megavoltage x-ray beam it uses for treatment. These megavoltage CT (MVCT) images offer verification of the patient position prior to and potentially during radiation therapy. Since the unit uses the actual treatment beam as the x-ray source for image acquisition, no surrogate telemetry systems are required to register image space to treatment space. The disadvantage to using the treatment beam for imaging, however, is that the physics of radiation interactions in the megavoltage energy range may force compromises between the dose delivered and the image quality in comparison to diagnostic CT scanners. The performance of the system is therefore characterized in terms of objective measures of noise, uniformity, contrast, and spatial resolution as a function of the dose delivered by the MVCT beam. The uniformity and spatial resolutions of MVCT images generated by the HI-ART II are comparable to that of diagnostic CT images. Furthermore, the MVCT scan contrast is linear with respect to the electron density of material imaged. MVCT images do not have the same performance characteristics as state-of-the art diagnostic CT scanners when one objectively examines noise and low-contrast resolution. These inferior results may be explained, at least partially, by the low doses delivered by our unit; the dose is 1.1 cGy in a 20 cm diameter cylindrical phantom. In spite of the poorer low-contrast resolution, these relatively low-dose MVCT scans provide sufficient contrast to delineate many soft-tissue structures. Hence, these images are useful not only for verifying the patient's position at the time of therapy, but they are also sufficient for delineating many anatomic structures. In conjunction with the ability to recalculate radiotherapy doses on these images, this enables dose guidance as well as image guidance of radiotherapy treatments.
螺旋断层放射治疗是一种采用结合了直线加速器和螺旋计算机断层扫描(CT)扫描仪功能的设备来进行图像引导放射治疗(IGRT)和调强放射治疗(IMRT)的创新方法。HI-ART II型设备能够利用其用于治疗的兆伏级X射线束生成CT图像。这些兆伏级CT(MVCT)图像可在放射治疗前及可能在治疗过程中对患者体位进行验证。由于该设备使用实际的治疗束作为图像采集的X射线源,因此无需替代遥测系统来将图像空间与治疗空间进行配准。然而,使用治疗束进行成像的缺点在于,与诊断CT扫描仪相比,兆伏级能量范围内的辐射相互作用物理特性可能会导致在剂量输送和图像质量之间做出妥协。因此,该系统的性能通过噪声、均匀性、对比度和空间分辨率等客观指标来表征,这些指标是兆伏级CT束所输送剂量的函数。HI-ART II型设备生成的MVCT图像的均匀性和空间分辨率与诊断CT图像相当。此外,MVCT扫描对比度与所成像材料的电子密度呈线性关系。当客观地检查噪声和低对比度分辨率时,MVCT图像并不具备与最先进的诊断CT扫描仪相同的性能特征。这些较差的结果至少部分可以由我们设备所输送的低剂量来解释;在直径为20厘米的圆柱形体模中,剂量为1.1厘戈瑞。尽管低对比度分辨率较差,但这些相对低剂量的MVCT扫描提供了足够的对比度来勾勒许多软组织结构。因此,这些图像不仅有助于在治疗时验证患者体位,而且也足以勾勒许多解剖结构。结合在这些图像上重新计算放射治疗剂量的能力,这使得能够进行剂量引导以及放射治疗的图像引导。