用于并行磁共振成像的同心线圈阵列。

Concentric coil arrays for parallel MRI.

作者信息

Ohliger Michael A, Greenman Robert L, Giaquinto Randy, McKenzie Charles A, Wiggins Graham, Sodickson Daniel K

机构信息

Harvard-MIT Division of Health Sciences and Technology, Boston, Massachusetts, USA.

出版信息

Magn Reson Med. 2005 Nov;54(5):1248-60. doi: 10.1002/mrm.20647.

Abstract

A new type of coil array is proposed that consists of concentrically placed coil elements, each of which is characterized by symmetrically arranged lobes that have alternating current directions. Symmetries in the coil elements' conductor paths allow for the minimization of mutual inductance and noise correlations. In addition, the concentric arrangement of the coil elements provides spatial encoding capabilities in multiple directions, which is valuable when arrays are used with parallel MRI. Simulations are presented that describe the signal-to-noise ratio (SNR) properties of individual concentric array elements, and a four-element prototype concentric array is constructed. This prototype array is compared experimentally with three alternative four-element array designs. The overall SNR of the concentric array is comparable to the SNR of the competing arrays. Reconstruction of twofold undersampled data using the concentric array yields an average g-factor of less than 1.3 in all directions parallel to the plane of the array. There is some degradation in performance when threefold undersampled data are reconstructed, but the array still shows substantial directional invariance compared to alternative designs. Both fully-sampled and undersampled cardiac images acquired using the concentric array are shown. These results suggest that concentric structures can be useful tools for designing specialized coil arrays for parallel MRI.

摘要

提出了一种新型线圈阵列,它由同心放置的线圈元件组成,每个线圈元件的特征是具有交替电流方向的对称排列的波瓣。线圈元件导体路径中的对称性使得互感和噪声相关性最小化。此外,线圈元件的同心排列提供了多方向的空间编码能力,这在阵列与并行磁共振成像(MRI)一起使用时很有价值。给出了描述单个同心阵列元件的信噪比(SNR)特性的模拟,并构建了一个四元件同心阵列原型。将该原型阵列与三种替代的四元件阵列设计进行了实验比较。同心阵列的整体SNR与竞争阵列的SNR相当。使用同心阵列重建两倍欠采样数据时,在与阵列平面平行的所有方向上,平均g因子小于1.3。当重建三倍欠采样数据时,性能会有一些下降,但与替代设计相比,该阵列仍显示出显著的方向不变性。展示了使用同心阵列采集的全采样和欠采样心脏图像。这些结果表明,同心结构可成为设计用于并行MRI的专用线圈阵列的有用工具。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验