Bocca Beatrice, Forte Giovanni, Petrucci Francesco, Senofonte Oreste, Violante Nicola, Alimonti Alessandro
Dipartimento di Ambiente e Connessa Prevenzione Primaria, Istituto Superiore di Sanità, Rome, Italy.
Ann Ist Super Sanita. 2005;41(2):165-70.
Analytical methods using sector field ICP-MS and ICP-AES were developed for the determination of Al, Ba, Be, Bi, Ca, Cd, Co, Cr, Cu, Fe, Hg, Li, Mg, Mn, Mo, Ni, Pb, Sb, Si, Sn, Sr, Tl, V, W, Zn and Zr in cerebrospinal fluid, urine, serum and blood. Sample treatment procedures merging high sample throughput, simplicity and low contamination risk were set up. Method performances were evaluated in terms of detection limits, accuracy and precision. The limits were below 0.05 ng ml(-1) for all the elements, except for Al (all matrices), Hg (blood), Pb (blood) and Sn (serum and blood). The accuracy varied from 86% to 110% and the precision was always below 6%.
开发了使用扇形磁场电感耦合等离子体质谱仪(ICP-MS)和电感耦合等离子体发射光谱仪(ICP-AES)的分析方法,用于测定脑脊液、尿液、血清和血液中的铝(Al)、钡(Ba)、铍(Be)、铋(Bi)、钙(Ca)、镉(Cd)、钴(Co)、铬(Cr)、铜(Cu)、铁(Fe)、汞(Hg)、锂(Li)、镁(Mg)、锰(Mn)、钼(Mo)、镍(Ni)、铅(Pb)、锑(Sb)、硅(Si)、锡(Sn)、锶(Sr)、铊(Tl)、钒(V)、钨(W)、锌(Zn)和锆(Zr)。建立了兼具高样品通量、操作简便和低污染风险的样品处理程序。从检测限、准确度和精密度方面对方法性能进行了评估。除了铝(所有基质)、汞(血液)、铅(血液)和锡(血清和血液)外,所有元素的检测限均低于0.05 ng ml⁻¹。准确度在86%至110%之间,精密度始终低于6%。