Stair ascent kinematics affect UHMWPE wear and damage in total knee replacements.

作者信息

Cottrell Jocelyn M, Babalola Omotunde, Furman Bridgette S, Wright Timothy M

机构信息

Laboratory for Biomedical Mechanics and Materials, Hospital for Special Surgery, New York, New York, USA.

出版信息

J Biomed Mater Res B Appl Biomater. 2006 Jul;78(1):15-9. doi: 10.1002/jbm.b.30451.

Abstract

Investigations were performed to characterize the wear patterns of tibial inserts in a load-controlled knee simulator by incorporating both normal gait and clinically acquired stair climbing kinematics and load conditions. Two different tibial insert designs were evaluated: the NexGen Cruciate Retaining Augmentable and NexGen Legacy Posterior Stabilized inserts. Two test conditions were run: standard gait only, and gait with bouts of stair climbing at a ratio of 70:1. Gravimetric wear measurements, damage and contact area assessments, and kinematic evaluations were performed. The addition of stair climbing kinematics significantly affected the wear behavior. Regardless of design, wear rates for standard gait tests were significantly higher than those that included bouts of stair climbing. The damage modes seen in both test conditions were primarily burnishing with secondary scratching and pitting. At 2 Mc, the damage areas were not significantly different between the two designs, but the damage area with stair climbing was significantly larger than that with gait alone. The fact that even small bouts of an additional activity of daily living could markedly impact wear simulator results calls into question the usefulness of studies that rely solely on kinematics and load inputs assumed from level gait.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验