Calibrated current divider network for precision current delivery during high-voltage transthoracic defibrillation.

作者信息

Deale O Carlton, Ng Kwong T, Lerman Bruce B

机构信息

Cardiology Division, Department of Medicine, Weill Medical College of Cornell University, New York, NY 10021, USA.

出版信息

IEEE Trans Biomed Eng. 2005 Nov;52(11):1970-3. doi: 10.1109/TBME.2005.856299.

Abstract

The design of a calibrated resistive-network current divider for precision current delivery during transthoracic defibrillation shocks is presented together with test results. The current divider presents a constant 50-ohm load to the defibrillator and thus maintains a constant pulse shape. Current is selected before the shock by setting three rheostats using a computer-generated calibration table. Following each shock, the data acquisition and display software updates the calibration table based on the measured value of transthoracic resistance. Over a range of 15-27 A, the root-mean-square (rms) error for delivered versus selected current was 0.48% for a 45-ohm resistive load, and 0.71% for a 100-ohm load. These test results were confirmed by animal experiments. Over 3 dogs, the rms error was 0.49% from 15-27 A and not greater than 1.5% over the entire 8-44 A range.