Kurki T J, Niemi P T, Lundbom N
Department of Diagnostic Radiology, Turku University Hospital, Finland.
J Magn Reson Imaging. 1992 Jul-Aug;2(4):401-6. doi: 10.1002/jmri.1880020408.
The magnetization transfer contrast (MTC) technique was used in low-field-strength (0.1 T) magnetic resonance (MR) imaging of 28 patients with intracranial tumors. MTC images were generated with an off-resonance, low-power radio-frequency pulse applied during the interpulse delay period of a gradient-echo partial-saturation sequence (TR msec/TE msec = 200/20). Images in the presence and absence of the MTC pulse were concurrently acquired before and after injection of gadopentetate dimeglumine at a dose of 0.1 mmol/kg. The contrast agent enhanced 27 of 28 tumors. Application of the MTC pulse improved the contrast-to-noise ratio (C/N) between tumor and normal white matter in 26 of 28 cases on the preinjection images and in 25 of 28 cases on the postinjection images. On the gadolinium-enhanced images, the mean C/N was 2.6 +/- 1.7 without the MTC pulse and 3.2 +/- 1.9 with the MTC pulse. The greatest contrast improvement with the MTC technique was obtained in tumors showing the strongest paramagnetic enhancement. The results indicate that MTC can improve contrast between normal brain and some intracranial neoplasms. The use of gadopentetate dimeglumine generally intensified this effect.
采用磁化传递对比(MTC)技术对28例颅内肿瘤患者进行低场强(0.1 T)磁共振(MR)成像。在梯度回波部分饱和序列(TR毫秒/TE毫秒 = 200/20)的脉冲间期施加偏离共振的低功率射频脉冲来生成MTC图像。在注射剂量为0.1 mmol/kg的钆喷酸葡胺之前和之后,同时采集有无MTC脉冲时的图像。造影剂使28个肿瘤中的27个增强。在注射前图像上,28例中有26例、注射后图像上28例中有25例,MTC脉冲的应用提高了肿瘤与正常白质之间的对比噪声比(C/N)。在钆增强图像上,无MTC脉冲时平均C/N为2.6±1.7,有MTC脉冲时为3.2±1.9。MTC技术在显示最强顺磁性增强的肿瘤中获得了最大的对比改善。结果表明,MTC可改善正常脑与某些颅内肿瘤之间的对比。钆喷酸葡胺的使用通常会增强这种效果。