Tulli G, Romagnoli P, Pratesi R, Vitale G G, Pacciani S, Giannoni S, Pappagallo S
U.O. Anestesia e Rianimazione, USL n. 18, Presidio Ospedaliero di Empoli, Firenze.
Minerva Anestesiol. 1992 May;58(5):289-95.
The Authors assess the utility of the on-line monitoring of SvO2 during 156 measurements of pulmonary capillary wedge pressure (PCWP) done on 52 Intensive Care Unit patients. The measurement was always right when the SvO2 increased more than 90%. The Authors found a good correlation between the SvO2 monitored during the measurement of the PCWP and the saturation of pulmonary capillary blood measured by cooximeter. These data suggest that the complex procedure to confirm the reliability of the measurement recommended by Gardner can be simplified by the observation of the trend of SvO2 on the monitor oximetrix. This allows to obtain saving time for the staff, saving blood for the patients, reduction of risk for the transmission of infective diseases due to the handling of blood and guarantees a further routinary control of reliability on the measure of PCWP.
作者评估了对52名重症监护病房患者进行156次肺毛细血管楔压(PCWP)测量期间对SvO2进行在线监测的效用。当SvO2增加超过90%时,测量结果总是正确的。作者发现,在测量PCWP期间监测的SvO2与通过血氧计测量的肺毛细血管血饱和度之间存在良好的相关性。这些数据表明,通过观察血氧饱和度监测仪上SvO2的趋势,可以简化加德纳推荐的用于确认测量可靠性的复杂程序。这为工作人员节省了时间,为患者节省了血液,降低了因处理血液而导致传染病传播的风险,并保证了对PCWP测量可靠性的进一步常规控制。