Lea Simon C, Landini Gabriel, Walmsley A Damien
School of Dentistry, The University of Birmingham, St. Chad's Queensway, Birmingham, UK.
J Clin Periodontol. 2006 Jan;33(1):37-41. doi: 10.1111/j.1600-051X.2005.00861.x.
During clinical usage, scaler tips may become worn and reduced in length. It is unknown what effect wear has on the magnitude of scaler tip vibrations when they are utilized under typical clinical loads. The aim of this investigation was to assess the effect of simulated wear on ultrasonic scaler tip displacement amplitude, using a scanning laser vibrometer.
A Cavitron SPS (Dentsply) ultrasonic generator and three scaler insert designs (FSI-100, FSI-1000 and FSI-SLI-10S) were selected for the investigation. Tip vibration displacement amplitude was assessed unloaded and then contacting against tooth surfaces with loads of 0.5 and 1.0 N. Tips were then ground down by 1 mm and then 2 mm and scans were repeated.
For all tips, load and length were found to be significant variables (p<0.0001). The scaler tips showed a fall in displacement amplitude with a reduction in tip length. However, all scaler tips showed variability in the amount oscillation that occurred. This was most pronounced with FSI-SLI-10S.
This investigation demonstrated that tip wear could affect the performance of dental ultrasonic scaler inserts by reducing their vibration displacement amplitude. Clinicians should be aware of this variability, which may be significant enough to affect clinical procedures.
在临床使用过程中,洁治器尖端可能会磨损并变短。在典型临床负荷下使用时,磨损对洁治器尖端振动幅度有何影响尚不清楚。本研究的目的是使用扫描激光振动计评估模拟磨损对超声洁治器尖端位移幅度的影响。
选择一台Cavitron SPS(登士柏)超声发生器和三种洁治器插入头设计(FSI - 100、FSI - 1000和FSI - SLI - 10S)进行研究。在无负载情况下评估尖端振动位移幅度,然后在0.5 N和1.0 N的负载下与牙齿表面接触进行评估。然后将尖端磨短1 mm,再磨短2 mm,并重复扫描。
对于所有尖端,负载和长度都是显著变量(p < 0.0001)。洁治器尖端的位移幅度随着尖端长度的缩短而下降。然而,所有洁治器尖端在发生的振荡量上都表现出变异性。这在FSI - SLI - 10S上最为明显。
本研究表明,尖端磨损会通过降低振动位移幅度影响牙科超声洁治器插入头的性能。临床医生应意识到这种变异性,其可能足以影响临床操作。