A polymer-based MR-compatible guidewire: a study to explore new prospects for interventional peripheral magnetic resonance angiography (ipMRA).

作者信息

Mekle Ralf, Hofmann Eugen, Scheffler Klaus, Bilecen Deniz

机构信息

MR-Physics, University of Basel/University Hospital, Basel, Switzerland.

出版信息

J Magn Reson Imaging. 2006 Feb;23(2):145-55. doi: 10.1002/jmri.20486.

Abstract

PURPOSE

To introduce a newly developed polymer-based and magnetic resonance (MR)-compatible guidewire and to explore its capabilities with respect to interventional peripheral magnetic resonance angiography (ipMRA) in a flow phantom.

MATERIALS AND METHODS

The guidewire is based on a polyetheretherketone (PEEK) polymer core, and small iron particles are embedded in its coating. A passive device tracking technique was designed utilizing a susceptibility artifact induced by the wire in images acquired with a balanced steady-state free precession (b-SSFP) sequence using small flip angles. The position of the guidewire tip was determined from image intensity maxima and overlayed onto a roadmap in near real-time. Guidewire tracking and balloon angioplasty of an artificial stenosis were attempted in two configurations of a flow phantom.

RESULTS

Successful passive guidewire tracking was performed for all phantom configurations. Robustness and accuracy of the tracking technique were sufficient for phantom studies. A balloon catheter was placed into the stenosis using the guidewire under complete MR guidance, and subsequent balloon angioplasty yielded improved flow conditions.

CONCLUSION

The new guidewire is well-suited for clinical application due to an absence of the risk of core fracture and its atraumatic flexible tip. It opens novel prospects for the realization of ipMRA in humans that need to be explored in further studies.