Batista E, Khoury H J, Melo F A, Barros V, da Silva E F
Departamento de Energia Nuclear, Universidade Federal de Pernambuco, Cidade Universitária, 50740-540 Recife, PE, Brazil.
Radiat Prot Dosimetry. 2005;115(1-4):391-3. doi: 10.1093/rpd/nci045.
This paper reports the results of an investigation carried out to determine the suitability of using a commercially available photodiode for dosimetry in mammography thereby providing a low cost dosemeter with a direct reading of the radiation dose. A mammographic X-ray generator (Siemens Mammomat 1000) with tube potential in the 23-30 kV range and a constant potential X-ray generator (Pantak) with a W/Mo anode/filter combination were used in this study. The results showed that the photodiode response is highly linear within mammographic dose ranges and that the energy dependence was <3% at tube potentials of 25-30 kV. Good agreement was observed between the incident air kerma measured with both the photodiode and the ionization chamber. These results show the viability of using the photodiode as a dosemeter system in mammography.
本文报告了一项调查结果,该调查旨在确定使用市售光电二极管进行乳腺摄影剂量测定的适用性,从而提供一种能够直接读取辐射剂量的低成本剂量计。本研究使用了管电压在23 - 30 kV范围内的乳腺X射线发生器(西门子Mammomat 1000)以及具有钨/钼阳极/滤过组合的恒压X射线发生器(潘塔克)。结果表明,在乳腺摄影剂量范围内,光电二极管的响应具有高度线性,并且在25 - 30 kV的管电压下,能量依赖性小于3%。用光电二极管和电离室测量的入射空气比释动能之间观察到了良好的一致性。这些结果表明了在乳腺摄影中使用光电二极管作为剂量计系统的可行性。