Suppr超能文献

[使用硒平板探测器的全视野数字化乳腺摄影中的辐射暴露]

[Radiation exposure in full-field digital mammography with a selenium flat-panel detector].

作者信息

Gosch D, Jendrass S, Scholz M, Kahn T

机构信息

Klinik und Poliklinik für Diagnostische und Interventionelle Radiologie, Universitätsklinikum Leipzig AöR.

出版信息

Rofo. 2006 Jul;178(7):693-7. doi: 10.1055/s-2006-926742. Epub 2006 Jun 7.

Abstract

PURPOSE

Calculation of the average glandular dose for mammography on a full-field digital mammography system using a selenium flat-panel detector.

MATERIALS AND METHODS

Mammographic examinations were carried out using the Selenia digital mammographic system from Lorad/Hologic. 1992 mammographies of 500 patients in cranio-caudal and medio-lateral projections were evaluated. Based on the recorded exposure conditions (tube voltage, tube loading, filtration, compressed breast thickness), the entrance surface air kerma was calculated by multiplying the tube loading by the measured tube output and was corrected according to the inverse square law. The average glandular dose was determined for each exposure by multiplying the entrance surface air kerma value by the relevant conversion factor for a breast composition of 50 % adipose tissue and 50 % glandular tissue by weight.

RESULTS

The mean values for patient age and compressed breast thickness were 61 years and 58 mm, respectively. The average glandular dose was 1.57 mGy for a single view (1.46 mGy for cranio-caudal view images and 1.68 mGy for medio-lateral view images).

CONCLUSION

Full-field digital mammography with a selenium flat-panel detector requires a dose similar to that of units with a flat-panel detector based on amorphous silicon and a dose approximately 20 % lower than that of conventional screen/film mammography.

摘要

目的

使用硒平板探测器的全视野数字乳腺摄影系统计算乳腺摄影的平均腺体剂量。

材料与方法

使用Lorad/Hologic公司的Selenia数字乳腺摄影系统进行乳腺摄影检查。对500例患者的1992次乳腺摄影的头尾位和内外侧斜位影像进行评估。根据记录的曝光条件(管电压、管负载、滤过、乳腺压迫厚度),通过将管负载乘以测量的管输出量来计算入射表面空气比释动能,并根据反平方定律进行校正。对于每次曝光,通过将入射表面空气比释动能值乘以重量组成为50%脂肪组织和50%腺体组织的乳腺的相关转换系数来确定平均腺体剂量。

结果

患者年龄和乳腺压迫厚度的平均值分别为61岁和58毫米。单视图的平均腺体剂量为1.57毫戈瑞(头尾位视图影像为1.46毫戈瑞,内外侧斜位视图影像为1.68毫戈瑞)。

结论

使用硒平板探测器的全视野数字乳腺摄影所需剂量与基于非晶硅的平板探测器设备相似,且比传统屏片乳腺摄影低约20%。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验