Werner Liliana, Hunter Brian, Stevens Scott, Chew Jesse J L, Mamalis Nick
John A. Moran Eye Center, University of Utah, 50 North Medical Drive, Salt Lake City, UT 84132, USA.
Am J Ophthalmol. 2006 Jan;141(1):35-43. doi: 10.1016/j.ajo.2005.08.045.
To verify the presence of the element silicon on hydrophilic acrylic intraocular lenses (IOLs) explanted because of calcification.
Interventional case series with clinicopathological correlation.
Twenty explanted IOLs with surface deposits (MemoryLens), and 20 with deposits mostly within their optic substance (SC60B-OUV and Aqua-Sense; 10 each) were used. After gross, microscopic, and histochemical analyses to confirm the presence of deposits, the lenses underwent scanning electron microscopy (SEM) with energy dispersive x-ray spectroscopy (EDS) for elemental composition, on the external surface of MemoryLens IOLs, and on the surface and internal substance of SC60B-OUV and Aqua-Sense IOLs. The weight percentage of the element silicon was obtained at the level of deposits, and at adjacent deposit-free areas in all lenses.
Scanning electron microscopy (SEM) coupled with EDS confirmed that the composition of the deposits was calcium/phosphate in all cases. The element silicon was found in all 40 lenses, on all areas analyzed. The silicon weight percentage was higher at the level of the deposits. The presence of aluminum on five MemoryLens IOLs, and in most of the SC60B-OUV and Aqua-Sense lenses might be related to scattering from the aluminum mounting stubs used for surface analyses.
Silicone compounds have been implicated in the calcification of another hydrophilic acrylic design (Hydroview). They may also have a role in the calcification of other hydrophilic acrylic IOLs. Further investigation on the relationship between the presence of the element silicon and the silicone compounds found on calcified hydrophilic acrylic lenses is necessary.
验证因钙化而被摘除的亲水性丙烯酸人工晶状体(IOL)上是否存在硅元素。
具有临床病理相关性的干预性病例系列研究。
使用了20枚表面有沉积物的被摘除IOL(MemoryLens),以及20枚沉积物大多位于其光学物质内部的IOL(SC60B - OUV和Aqua - Sense,各10枚)。在进行大体、显微镜和组织化学分析以确认沉积物的存在后,对这些晶状体进行扫描电子显微镜(SEM)检查,并结合能量色散X射线光谱(EDS)分析其元素组成,分析部位包括MemoryLens IOL的外表面,以及SC60B - OUV和Aqua - Sense IOL的表面和内部物质。在所有晶状体的沉积物部位以及相邻的无沉积物区域获取硅元素的重量百分比。
扫描电子显微镜(SEM)结合EDS证实,所有病例中沉积物的成分均为钙/磷酸盐。在所有分析区域的所有40枚晶状体中均发现了硅元素。沉积物部位的硅重量百分比更高。在5枚MemoryLens IOL以及大多数SC60B - OUV和Aqua - Sense晶状体上发现的铝元素,可能与用于表面分析的铝制安装短柱的散射有关。
有机硅化合物与另一种亲水性丙烯酸设计(Hydroview)的钙化有关。它们也可能在其他亲水性丙烯酸IOL的钙化过程中起作用。有必要进一步研究硅元素的存在与钙化亲水性丙烯酸晶状体上发现的有机硅化合物之间的关系。