Image overlay for CT-guided needle insertions.

作者信息

Fichtinger Gabor, Deguet Anton, Fischer Gregory, Iordachita Iulian, Balogh Emese, Masamune Ken, Taylor Russell H, Fayad Laura M, de Oliveira Michelle, Zinreich S James

机构信息

Johns Hopkins University, Baltimore, MD 21218-2682, USA.

出版信息

Comput Aided Surg. 2005 Jul;10(4):241-55. doi: 10.3109/10929080500230486.

Abstract

OBJECTIVE

We present a 2D image overlay device to assist needle placement on computed tomography (CT) scanners.

MATERIALS AND METHODS

The system consists of a flat display and a semitransparent mirror mounted on the gantry. When the physician looks at the patient through the mirror, the CT image appears to be floating inside the body with correct size and position as if the physician had 2D 'X-ray vision'. The physician draws the optimal path on the CT image. The composite image is rendered on the display and thus reflected in the mirror. The reflected image is used to guide the physician in the procedure. In this article, we describe the design and various embodiments of the 2D image overlay system, followed by the results of phantom and cadaver experiments in multiple clinical applications.

RESULTS

Multiple skeletal targets were successfully accessed with one insertion attempt. Generally, successful access was recorded on liver targets when a clear path opened, but the number of attempts and accuracy showed variability because of occasional lack of access. Soft tissue deformation further reduced the accuracy and consistency in comparison to skeletal targets.

CONCLUSION

The system demonstrated strong potential for reducing faulty needle insertion attempts, thereby reducing X-ray dose and patient discomfort.