J Pharm Biomed Anal. 2006 May 3;41(2):527-32. doi: 10.1016/j.jpba.2005.11.021. Epub 2006 Jan 6.

Abstract

An isocratic high-performance liquid chromatographic method with detection at 234 nm was developed, optimized and validated for the determination of testosterone in human serum. Propylparaben was used as internal standard. A Hypersil BDS RP-C18 column (150 mmx4.6 mm, 5 microm), was equilibrated with a mobile phase composed of acetonitrile and water (35:65, v/v) and having a flow rate of 1 ml/min. The elution time for testosterone and internal standard was approximately 11.6 and 9.9 min, respectively. Calibration curves of testosterone in serum were linear in the concentration range of 1-20 ng/ml. Limits of detection and quantification in serum were 0.4 and 1.1 ng/ml, respectively. Recovery was greater than 92%. Intra- and inter-day relative standard deviation for testosterone in serum was less than 2.1 and 3.9%, respectively. This method was applied to the determination of testosterone serum levels of 12 healthy males and data were correlated with data obtained using a radioimmunoassay method.

摘要

建立了一种在234nm波长下检测的等度高效液相色谱法,用于测定人血清中的睾酮,并对其进行了优化和验证。对羟基苯甲酸丙酯用作内标。使用Hypersil BDS RP-C18柱(150mm×4.6mm,5μm),用由乙腈和水(35:65,v/v)组成的流动相平衡,流速为1ml/min。睾酮和内标的洗脱时间分别约为11.6和9.9分钟。血清中睾酮的校准曲线在1-20ng/ml的浓度范围内呈线性。血清中的检测限和定量限分别为0.4和1.1ng/ml。回收率大于92%。血清中睾酮的日内和日间相对标准偏差分别小于2.1%和3.9%。该方法应用于12名健康男性血清睾酮水平的测定,并将数据与使用放射免疫分析法获得的数据进行了关联。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验