Eccles Cynthia, Brock Kristy K, Bissonnette Jean-Pierre, Hawkins Maria, Dawson Laura A
Radiation Medicine Program, Princess Margaret Hospital, University of Toronto, Toronto, Ontario, Canada.
Int J Radiat Oncol Biol Phys. 2006 Mar 1;64(3):751-9. doi: 10.1016/j.ijrobp.2005.05.066.
To measure the intrabreath-hold liver motion and the intrafraction and interfraction reproducibility of liver position relative to vertebral bodies using an active breathing coordinator (ABC) in patients with unresectable liver cancer treated with hypofractionated stereotactic body radiation therapy (SBRT).
Tolerability of ABC and organ motion during ABC was assessed using kV fluoroscopy in 34 patients. For patients treated with ABC, repeat breath-hold CT scans in the ABC breath-hold position were acquired at simulation to estimate the volumetric intrafraction reproducibility of the liver relative to the vertebral bodies. In addition, preceding each radiation therapy fraction, with the liver immobilized using ABC, repeat anteroposterior (AP) megavoltage verification images were obtained. Off-line alignments were completed to determine intrafraction reproducibility (from repeat images obtained before one treatment) and interfraction reproducibility (from comparisons of the final image for each fraction with the AP) of diaphragm position relative to vertebral bodies. For each image set, the vertebral bodies were aligned, and the resultant craniocaudal (CC) offset in diaphragm position was measured. Liver position during ABC was also evaluated from kV fluoroscopy acquired at the time of simulation, kV fluoroscopy at the time of treatment, and from MV beam's-eye view movie loops acquired during treatment.
Twenty-one of 34 patients were screened to be suitable for ABC. The average free breathing range of these patients was 13 mm (range, 5-1 mm). Fluoroscopy revealed that the average maximal diaphragm motion during ABC breath-hold was 1.4 mm (range, 0-3.4 mm). The MV treatment movie loops confirmed diaphragm stability during treatment. For a measure of intrafraction reproducibility, an analysis of 36 repeat ABC computed tomography (CT) scans in 14 patients was conducted. The average mean difference in the liver surface position was -0.9 mm, -0.5 mm, and 0.2 mm in the CC, AP, and medial-lateral (ML) directions, with a standard deviation of 1.5 mm, 1.5 mm, and 1.5 mm, respectively. Ninety-five percent of the liver surface had an absolute differences in position between repeat ABC CT scans of less than 4.1 mm, 3.3 mm, and 3.3 mm in the CC, AP, and ML directions, respectively. Analysis of 257 MV AP images from patients treated using ABC revealed an average intrafraction CC reproducibility (sigma) of diaphragm relative to vertebral bodies of 1.5 mm (range, 0.6-3.9 mm). The average interfraction CC reproducibility (sigma) was 3.4 mm (range, 1.5-7.9 mm), indicating less day-to-day reproducibility of diaphragm position relative to vertebral bodies. The average absolute intra and interfraction CC offset in diaphragm position relative to vertebral bodies was 1.7 and 3.7 mm, respectively, with 86% of intrafraction and 54% of interfraction absolute offsets 3.0 mm or less.
Intrafraction reproducibility of liver position using ABC is good in the majority of screened patients. However, interfraction reproducibility is worse, suggesting a need for image guidance.
使用主动呼吸控制仪(ABC)测量不可切除肝癌患者在分次立体定向体部放射治疗(SBRT)过程中屏气时肝脏的运动以及肝脏相对于椎体位置的分次内和分次间可重复性。
在34例患者中使用千伏透视评估ABC的耐受性以及ABC期间的器官运动。对于接受ABC治疗的患者,在模拟时获取ABC屏气位置的重复屏气CT扫描,以估计肝脏相对于椎体的容积分次内可重复性。此外,在每次放射治疗分次前,使用ABC固定肝脏,获取重复的前后位(AP)兆伏级验证图像。完成离线配准以确定膈肌相对于椎体位置的分次内可重复性(从一次治疗前获得的重复图像)和分次间可重复性(从每个分次的最终图像与AP的比较)。对于每组图像,对齐椎体,并测量膈肌位置产生的头脚向(CC)偏移。还从模拟时获取的千伏透视、治疗时的千伏透视以及治疗期间获取的兆伏级射野视角电影环中评估ABC期间的肝脏位置。
34例患者中有21例被筛选适合使用ABC。这些患者的平均自由呼吸范围为13毫米(范围为5 - 1毫米)。透视显示ABC屏气期间膈肌的平均最大运动为1.4毫米(范围为0 - 3.4毫米)。兆伏级治疗电影环证实了治疗期间膈肌的稳定性。对于分次内可重复性的测量,对14例患者的36次重复ABC计算机断层扫描(CT)进行了分析。肝脏表面位置在CC、AP和内外侧(ML)方向的平均平均差异分别为 - 0.9毫米、 - 0.5毫米和0.2毫米,标准差分别为1.5毫米、1.5毫米和1.5毫米。95%的肝脏表面在重复ABC CT扫描时CC、AP和ML方向的位置绝对差异分别小于4.1毫米、3.3毫米和3.3毫米。对使用ABC治疗的患者的257张MV AP图像的分析显示,膈肌相对于椎体的平均分次内CC可重复性(sigma)为1.5毫米(范围为0.6 - 3.9毫米)。平均分次间CC可重复性(sigma)为3.4毫米(范围为1.5 - 7.9毫米),表明膈肌相对于椎体位置的每日可重复性较差。膈肌相对于椎体位置的平均绝对分次内和分次间CC偏移分别为1.7毫米和3.7毫米,86%的分次内和54%的分次间绝对偏移为3.0毫米或更小。
在大多数筛选出的患者中,使用ABC时肝脏位置的分次内可重复性良好。然而,分次间可重复性较差,提示需要图像引导。