Novel breath-hold liver target stereotactic ablative radiotherapy using the intrafraction diaphragm registration of kilovoltage projection streaming image with digitally reconstructed radiography of the planning computed tomography.

作者信息

Katano Atsuto, Nozawa Yuki, Minamitani Masanari, Yamashita Hideomi, Nakagawa Keiichi

机构信息

Department of Radiology, The University of Tokyo Hospital, 7-3-1 Hongo, Bunkyo-ku, Tokyo 113-8655, Japan.

Department of Comprehensive Radiation Oncology, The University of Tokyo, 7-3-1 Hongo Bunkyo-ku, Tokyo 113-8655, Japan.

出版信息

Tech Innov Patient Support Radiat Oncol. 2023 Jun 28;27:100217. doi: 10.1016/j.tipsro.2023.100217. eCollection 2023 Sep.

Abstract

Stereotactic ablative radiotherapy (SABR) is an emerging treatment option for patients with primary or metastatic liver tumors, particularly for those who are not eligible for surgery or transplantation. SABR is a high-precision radiation therapy that delivers a high dose of radiation to the tumor while minimizing the dose to the surrounding healthy tissues. However, the accurate targeting of the tumor is a crucial aspect of liver SABR, which requires real-time imaging and tracking of the liver and tumor motion during treatment. One of the motion management strategies for liver SABR is the repeated breath-hold technique, which involves the patient holding their breath multiple times during treatment delivery to reduce the movement of the liver and other organs due to breathing. This technique helps to improve the accuracy of the treatment and reduce the radiation dose to the healthy liver. The current study proposes a novel approach for multiple breath-hold volumetric modulated arc therapy (VMAT) stereotactic ablative radiotherapy for liver tumors, which uses the intrafraction diaphragm registration in real time to improve the accuracy and precision of the treatment. The proposed approach is based on real-time comparison of the diaphragmatic surface location between the digitally reconstructed radiography (DRR) and intrafraction kilovoltage projection streaming images (kV-PSI) having the same beam angles. The image cross-correlation between the DRR and the intrafraction kV-PSI provides a measure of the similarity between the two images and can be used to identify and track the diaphragm position during VMAT delivery. The proposed methodology consists of several steps, including planning CT and treatment planning, reference image reconstruction, and patient positioning and immobilization. The proposed approach has the potential to improve the accuracy and precision of liver cancer VMAT SABR, thereby increasing the efficacy of the treatment and reducing the risk of radiation exposure to surrounding healthy tissues.

摘要

立体定向消融放疗(SABR)是一种针对原发性或转移性肝肿瘤患者的新兴治疗选择,尤其适用于那些不适合手术或移植的患者。SABR是一种高精度放疗,可在将高剂量辐射传递至肿瘤的同时,将对周围健康组织的剂量降至最低。然而,准确靶向肿瘤是肝脏SABR的一个关键方面,这需要在治疗过程中对肝脏和肿瘤运动进行实时成像和跟踪。肝脏SABR的运动管理策略之一是重复屏气技术,即在治疗过程中患者多次屏气,以减少由于呼吸导致的肝脏和其他器官的运动。该技术有助于提高治疗的准确性,并降低对健康肝脏的辐射剂量。当前研究提出了一种用于肝脏肿瘤的多次屏气容积调强弧形放疗(VMAT)立体定向消融放疗的新方法,该方法实时使用分次内膈肌配准来提高治疗的准确性和精确性。所提出的方法基于对具有相同射束角度的数字重建射线照相(DRR)和分次内千伏投影流图像(kV-PSI)之间膈肌表面位置的实时比较。DRR与分次内kV-PSI之间的图像互相关提供了两种图像之间相似性的度量,并可用于在VMAT治疗过程中识别和跟踪膈肌位置。所提出的方法包括几个步骤,包括计划CT和治疗计划、参考图像重建以及患者定位和固定。所提出的方法有可能提高肝癌VMAT SABR的准确性和精确性,从而提高治疗效果并降低对周围健康组织的辐射暴露风险。