Doyle P, Finney L
Department of Clinical Physics and Bio-Engineering, Gartnavel Royal Hospital, Glasgow G12 OXH, UK.
Radiat Prot Dosimetry. 2005;117(1-3):313-7. doi: 10.1093/rpd/nci762. Epub 2006 Feb 3.
Four digital intra-oral radiographic systems were tested and evaluated; three charge-coupled device (CCD) based systems from RVG, Visualix and Sidexis and a photostimulable phosphor (PSP) system from DenOptix. Image quality was assessed using three purpose-built phantoms to measure uniformity, low contrast detail detectability and signal-to-noise ratio (SNR). Limiting resolution was measured using a 20 lp mm(-1) bar pattern. Radiation output inaccuracies caused difficulties in achieving optimum doses for CCD systems. However, the accuracy was improved by using K-edge filters. SNR measurements proved to be a useful tool in assessing system performance. Each system has specific attributes: resolution was highest for the RVG system, the Visualix system measured the highest SNR and the lowest exposure settings were on the Sidexis system. Test methods and phantoms developed are suitable for acceptance testing and commissioning digital dental X-ray systems and for programming each system to produce an optimum level of image quality.
对四种数字化口腔内放射成像系统进行了测试和评估;三种基于电荷耦合器件(CCD)的系统,分别来自RVG、Visualix和Sidexis,以及一种来自DenOptix的光激励荧光体(PSP)系统。使用三个特制模体评估图像质量,以测量均匀性、低对比度细节可探测性和信噪比(SNR)。使用20 lp mm(-1) 的条形图案测量极限分辨率。辐射输出不准确给CCD系统实现最佳剂量带来困难。然而,通过使用K边滤波器提高了准确性。信噪比测量被证明是评估系统性能的有用工具。每个系统都有特定属性:RVG系统的分辨率最高,Visualix系统测量的信噪比最高,Sidexis系统的曝光设置最低。所开发的测试方法和模体适用于数字牙科X射线系统的验收测试和调试,以及对每个系统进行编程以产生最佳水平的图像质量。