不同激发磷光板擦除次数对主观数字图像质量的影响。

Effect of alternative photostimulable phosphor plates erasing times on subjective digital image quality.

机构信息

Department of Oral Diagnosis, Division of Oral Radiology, Piracaba Dental School, Campinas University, UNICAMP, Piracaba, Brazil.

出版信息

Dentomaxillofac Radiol. 2010 Jan;39(1):23-7. doi: 10.1259/dmfr/49065239.

Abstract

OBJECTIVE

To evaluate the influence of alternative erasing times of DenOptix(R) (Dentsply/Gendex, Chicargo, IL) digital plates on subjective image quality and the probability of double exposure image not occurring.

METHODS

Human teeth were X-rayed with phosphor plates using ten different erasing times. Two observers evaluated the images for subjective image quality (sharpness, brightness, contrast, enamel definition, dentin definition and dentin-enamel junction definition) and for the presence or absence of double exposure image. Spearman's correlation analysis and ANOVA was performed to verify the existence of a linear association between the subjective image quality parameters and the alternative erasing times. A contingency table was constructed to evaluate the agreement among the observers, and a binominal logistic regression was performed to verify the correlation between the erasing time and the probability of double exposure image not occurring.

RESULTS

All 6 parameters of image quality were rated high by the examiners for the erasing times between 25 s and 130 s. The same erasing time range, from 25 to 130 s, was considered a safe erasing time interval, with no probability of a double exposure image occurring.

CONCLUSIONS

The alternative erasing times from 25 s to 130 s showed high image quality and no probability of double image occurrence. Thus, it is possible to reduce the operating time of the DenOptix(R) digital system without jeopardizing the diagnostic task.

摘要

目的

评估 DenOptix(R)(登士柏/根德,芝加哥,IL)数字板的替代擦除次数对主观图像质量和双曝光图像发生概率的影响。

方法

使用十种不同的擦除时间对人牙进行磷光板 X 射线照相。两名观察者评估图像的主观图像质量(清晰度、亮度、对比度、釉质定义、牙本质定义和牙本质-釉质交界处定义)和双曝光图像的存在与否。采用 Spearman 相关分析和方差分析来验证主观图像质量参数与替代擦除时间之间是否存在线性关联。构建列联表来评估观察者之间的一致性,并进行二项逻辑回归来验证擦除时间与双曝光图像不发生概率之间的相关性。

结果

在 25 秒至 130 秒的擦除时间范围内,所有 6 项图像质量参数均被检查者评为高分。从 25 秒到 130 秒的相同擦除时间范围被认为是安全的擦除时间间隔,没有发生双曝光图像的概率。

结论

替代擦除时间为 25 秒至 130 秒,图像质量高,且无双图像发生的概率。因此,可以在不影响诊断任务的情况下减少 DenOptix(R)数字系统的操作时间。

相似文献

1
Effect of alternative photostimulable phosphor plates erasing times on subjective digital image quality.
Dentomaxillofac Radiol. 2010 Jan;39(1):23-7. doi: 10.1259/dmfr/49065239.
2
Alternative erasing times of the DenOptix system plate: performance on the detection of proximal caries.
Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2009 Jan;107(1):122-6. doi: 10.1016/j.tripleo.2008.04.033. Epub 2008 Jul 25.
3
Image quality in partially erased DenOptix storage phosphor plates.
Braz Oral Res. 2008 Jan-Mar;22(1):78-83. doi: 10.1590/s1806-83242008000100014.
5
Accuracy of caries detection with four storage phosphor systems and E-speed radiographs.
Dentomaxillofac Radiol. 2002 May;31(3):170-5. doi: 10.1038/sj/dmfr/4600686.
6
Comparison of three intra-oral storage phosphor systems using subjective image quality.
Dentomaxillofac Radiol. 2000 Sep;29(5):272-6. doi: 10.1038/sj/dmfr/4600532.
7
Reproducibility of pixel values for two photostimulable phosphor plates in consecutive standardized scannings.
Braz Oral Res. 2006 Jul-Sep;20(3):207-13. doi: 10.1590/s1806-83242006000300005.
10

引用本文的文献

1
Evaluation of the Effect of Light and Scanning Time Delay on The Image Quality of Intra Oral Photostimulable Phosphor Plates.
Open Dent J. 2017 Dec 29;11:690-700. doi: 10.2174/1874210601711010690. eCollection 2017.
2
Effect of Different Tube Potential Settings on Caries Detection using PSP Plate and Conventional Film.
J Clin Diagn Res. 2015 Apr;9(4):ZC58-61. doi: 10.7860/JCDR/2015/12225.5845. Epub 2015 Apr 1.
3
The effect of delayed scanning of storage phosphor plates on occlusal caries detection.
Dentomaxillofac Radiol. 2012 May;41(4):309-15. doi: 10.1259/dmfr/12935491. Epub 2012 Jan 26.

本文引用的文献

1
Alternative erasing times of the DenOptix system plate: performance on the detection of proximal caries.
Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2009 Jan;107(1):122-6. doi: 10.1016/j.tripleo.2008.04.033. Epub 2008 Jul 25.
2
Image quality in partially erased DenOptix storage phosphor plates.
Braz Oral Res. 2008 Jan-Mar;22(1):78-83. doi: 10.1590/s1806-83242008000100014.
3
Time and motion study: a comparison of two photostimulable phosphor imaging systems used in dentistry.
Dentomaxillofac Radiol. 2006 Sep;35(5):315-8. doi: 10.1259/dmfr/29518441.
4
Performance evaluation and testing of digital intra-oral radiographic systems.
Radiat Prot Dosimetry. 2005;117(1-3):313-7. doi: 10.1093/rpd/nci762. Epub 2006 Feb 3.
6
The dynamic range of digital radiographic systems: dose reduction or risk of overexposure?
Dentomaxillofac Radiol. 2004 Jan;33(1):1-5. doi: 10.1259/dmfr/40677472.
9
Comments on noise and resolution of the DenOptix radiography system.
Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2003 Jun;95(6):746-51. doi: 10.1067/moe.2003.245.
10
Accuracy of computer-automated caries detection in digital radiographs compared with human observers.
Eur J Oral Sci. 2002 Jun;110(3):199-203. doi: 10.1034/j.1600-0447.2002.21245.x.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验