Dentomaxillofac Radiol. 2010 Jan;39(1):23-7. doi: 10.1259/dmfr/49065239.
To evaluate the influence of alternative erasing times of DenOptix(R) (Dentsply/Gendex, Chicargo, IL) digital plates on subjective image quality and the probability of double exposure image not occurring.
Human teeth were X-rayed with phosphor plates using ten different erasing times. Two observers evaluated the images for subjective image quality (sharpness, brightness, contrast, enamel definition, dentin definition and dentin-enamel junction definition) and for the presence or absence of double exposure image. Spearman's correlation analysis and ANOVA was performed to verify the existence of a linear association between the subjective image quality parameters and the alternative erasing times. A contingency table was constructed to evaluate the agreement among the observers, and a binominal logistic regression was performed to verify the correlation between the erasing time and the probability of double exposure image not occurring.
All 6 parameters of image quality were rated high by the examiners for the erasing times between 25 s and 130 s. The same erasing time range, from 25 to 130 s, was considered a safe erasing time interval, with no probability of a double exposure image occurring.
The alternative erasing times from 25 s to 130 s showed high image quality and no probability of double image occurrence. Thus, it is possible to reduce the operating time of the DenOptix(R) digital system without jeopardizing the diagnostic task.
评估 DenOptix(R)(登士柏/根德,芝加哥,IL)数字板的替代擦除次数对主观图像质量和双曝光图像发生概率的影响。
使用十种不同的擦除时间对人牙进行磷光板 X 射线照相。两名观察者评估图像的主观图像质量(清晰度、亮度、对比度、釉质定义、牙本质定义和牙本质-釉质交界处定义)和双曝光图像的存在与否。采用 Spearman 相关分析和方差分析来验证主观图像质量参数与替代擦除时间之间是否存在线性关联。构建列联表来评估观察者之间的一致性,并进行二项逻辑回归来验证擦除时间与双曝光图像不发生概率之间的相关性。
在 25 秒至 130 秒的擦除时间范围内,所有 6 项图像质量参数均被检查者评为高分。从 25 秒到 130 秒的相同擦除时间范围被认为是安全的擦除时间间隔,没有发生双曝光图像的概率。
替代擦除时间为 25 秒至 130 秒,图像质量高,且无双图像发生的概率。因此,可以在不影响诊断任务的情况下减少 DenOptix(R)数字系统的操作时间。