数字减影血管造影术给患者带来的辐射剂量。

Radiation doses to patients from digital subtraction angiography.

作者信息

Goni H, Tsalafoutas I A, Tzortzis G, Pappas P, Bouzas N, Loulakas J, Georgiou A, Georgiou E, Yakoumakis E N

机构信息

Medical Physics Unit, Laiko Hospital, 17 Ag. Thoma, Goudi, 11527 Athens, Greece.

出版信息

Radiat Prot Dosimetry. 2005;117(1-3):251-5. doi: 10.1093/rpd/nci758. Epub 2006 Feb 3.

Abstract

This investigation determined patient doses during digital subtraction angiography (DSA). Fluoroscopy time, dose-area product (DAP) and entrance surface air kerma (ESAK) were analysed from 263 DSA examinations, classified into seven categories: (1) abdominal aorta, iliac, femoral, popliteal and leg arteries; (2) abdominal aorta and superselective DSA of renal arteries; (3) combination of (1) and (2); (4) superselective DSA of common carotid and vertebral arteries, intracranial branches in face and profile projections; (5) superselective DSA of hepatic, splenic, superior and inferior mesenteric arteries; (6) combination of (1) and (4); and (7) celiac trunk and branches. Median DAP values were 67.7, 92.9, 76.6, 53.6, 105.7, 76.1 and 2.6 Gy cm2, respectively. With the exception of one examination, ESAK values were below 2 Gy: the limit for erythema. Compared with published data, DAP values were within the range reported for (1) and (4), slightly larger for (2) and (5), whereas no references were identified for the remaining three categories.

摘要

本研究测定了数字减影血管造影(DSA)期间的患者剂量。对263例DSA检查的透视时间、剂量面积乘积(DAP)和体表空气比释动能(ESAK)进行了分析,这些检查分为七类:(1)腹主动脉、髂动脉、股动脉、腘动脉和腿部动脉;(2)腹主动脉和肾动脉超选择性DSA;(3)(1)和(2)的组合;(4)颈总动脉和椎动脉超选择性DSA,面部和侧位投影的颅内分支;(5)肝动脉、脾动脉、肠系膜上动脉和肠系膜下动脉超选择性DSA;(6)(1)和(4)的组合;(7)腹腔干及其分支。DAP中位数分别为67.7、92.9、76.6、53.6、105.7、76.1和2.6 Gy cm²。除一次检查外,ESAK值均低于2 Gy(红斑剂量限值)。与已发表的数据相比,DAP值在(1)和(4)报告的范围内,(2)和(5)略大,而其余三类未找到相关参考数据。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验