Haraldsson P, Karlsson A, Wieslander E, Gustavsson H, Bäck S A J
Medical Radiation Physics, Department of Clinical Sciences, Lund University, Malmö University Hospital, SE-205 02 Malmö, Sweden.
Phys Med Biol. 2006 Feb 21;51(4):919-28. doi: 10.1088/0031-9155/51/4/011. Epub 2006 Feb 1.
A low-density (approximately 0.6 g cm(-3)) normoxic polymer gel, containing the antioxidant tetrakis (hydroxymethyl) phosponium (THP), has been investigated with respect to basic absorbed dose response characteristics. The low density was obtained by mixing the gel with expanded polystyrene spheres. The depth dose data for 6 and 18 MV photons were compared with Monte Carlo calculations. A large volume phantom was irradiated in order to study the 3D dose distribution from a 6 MV field. Evaluation of the gel was carried out using magnetic resonance imaging. An approximately linear response was obtained for 1/T2 versus dose in the dose range of 2 to 8 Gy. A small decrease in the dose response was observed for increasing concentrations of THP. A good agreement between measured and Monte Carlo calculated data was obtained, both for test tubes and the larger 3D phantom. It was shown that a normoxic polymer gel with a reduced density could be obtained by adding expanded polystyrene spheres. In order to get reliable results, it is very important to have a uniform distribution of the gel and expanded polystyrene spheres in the phantom volume.
一种含有抗氧化剂四(羟甲基)鏻(THP)的低密度(约0.6 g cm⁻³)常氧聚合物凝胶,已针对基本吸收剂量响应特性进行了研究。通过将凝胶与膨胀聚苯乙烯球混合获得低密度。将6和18 MV光子的深度剂量数据与蒙特卡罗计算结果进行了比较。为了研究6 MV射野的三维剂量分布,对一个大体积模体进行了照射。使用磁共振成像对凝胶进行评估。在2至8 Gy的剂量范围内,1/T2与剂量之间获得了近似线性响应。随着THP浓度增加,观察到剂量响应有小幅下降。对于试管和较大的三维模体,测量数据与蒙特卡罗计算数据均取得了良好的一致性。结果表明,通过添加膨胀聚苯乙烯球可以获得密度降低的常氧聚合物凝胶。为了获得可靠的结果,使凝胶和膨胀聚苯乙烯球在模体体积内均匀分布非常重要。