Karaiskos P, Petrokokkinos L, Tatsis E, Angelopoulos A, Baras P, Kozicki M, Papagiannis P, Rosiak J M, Sakelliou L, Sandilos P, Vlachos L
Medical Physics Department, Hygeia Hospital, Kiffisias Avenue, 4 Erythrou Stavrou, Marousi, 151 23 Athens, Greece.
Phys Med Biol. 2005 Mar 21;50(6):1235-50. doi: 10.1088/0031-9155/50/6/013. Epub 2005 Mar 2.
This work describes an experimental procedure with potential to assess the overall accuracy associated with gamma knife clinical applications, from patient imaging and dosimetry planning to patient positioning and dose delivery using the automated positioning system of a Leksell Gamma Knife model C. The VIPAR polymer gel-MRI dosimetry method is employed due to its inherent three-dimensional feature and linear dose response over the range of gamma knife applications. Different polymer gel vials were irradiated with single shot gamma knife treatment plans using each of the four available collimator helmets to deliver a maximum dose of 30 Gy. Percentage relative dose results are presented not only in the form of one-dimensional profiles but also planar isocontours and isosurfaces in three dimensions. Experimental results are compared with corresponding Gammaplan treatment planning system calculations as well as acceptance test radiochromic film measurements. A good agreement, within the experimental uncertainty, is observed between measured and expected dose distributions. This experimental uncertainty is of the order of one imaging pixel in the MRI gel readout session (<1 mm) and allows for the verification of single shot gamma knife applications in terms of acceptance specifications for precision in beam alignment and accuracy. Averaging net R(2) results in the dose plateau of the 4 mm and 18 mm collimator irradiated gel vials, which were MR scanned in the same session, provides a crude estimate of the 4 mm output factor which agrees within errors with the default value of 0.870.
这项工作描述了一种实验程序,该程序有潜力评估与伽玛刀临床应用相关的整体准确性,涵盖从患者成像和剂量测定规划到使用Leksell伽玛刀C型自动定位系统进行患者定位和剂量输送的全过程。采用VIPAR聚合物凝胶-MRI剂量测定法是因为其固有的三维特性以及在伽玛刀应用范围内的线性剂量响应。使用四个可用的准直器头盔中的每一个,通过单次伽玛刀治疗计划对不同的聚合物凝胶瓶进行照射,以提供30 Gy的最大剂量。相对剂量百分比结果不仅以一维剖面图的形式呈现,还以三维平面等剂量线和等剂量面的形式呈现。将实验结果与相应的伽玛计划治疗计划系统计算结果以及验收测试放射变色胶片测量结果进行比较。在实验不确定度范围内,观察到测量剂量分布与预期剂量分布之间有良好的一致性。这种实验不确定度在MRI凝胶读出过程中约为一个成像像素(<1 mm),并允许根据光束对准精度和准确性的验收规范来验证单次伽玛刀应用。对在同一次扫描中进行MR扫描的4 mm和18 mm准直器照射的凝胶瓶剂量平台中的净R(2)结果进行平均,可对4 mm输出因子进行粗略估计,该估计值在误差范围内与默认值0.870一致。