Cameron D, Arnot R, Clay M, Silverman M
Department of Pediatrics and Neonatal Medicine, Royal Postgraduate Medical School, Hammersmith Hospital, London, England.
Pediatr Pulmonol. 1991;10(3):208-13. doi: 10.1002/ppul.1950100314.
The benefits of inhaled therapy in ventilated neonates are recognized, but the reliability of drug delivery in nebulizer-ventilator circuits is uncertain. We quantified the effect of changing variables. Twenty-three freshly killed rabbits (1.15-1.9 kg) were ventilated via a tracheostomy by a pressure-limited, time-cycled ventilator (Neovent). A radioaerosol of 99Tcm pertechnetate from an Ultravent nebulizer (Mallinkrodt) was fed into the proximal ventilator tubing. Two 3-minute nebulizations at "standard settings" were followed by 2 at altered pressure, frequency, gas flow, I:E ratio, or position of the nebulizer in the circuit. Each nebulization was followed by a 3-minute gamma camera image and total deposited radioactivity was measured in excised lungs and trachea. Images demonstrated good peripheral aerosol deposition. At standard settings, lung deposition averaged 2.8% of the aerosol released. This was decreased markedly by reducing tidal volume (ventilator pressures) and residence time of aerosol (I:E ratio). Reduced gas flow decreased deposition slightly, presumably by increased particle size and marginally reduced tidal volume. Deposition did not change with increased frequency; increased minute ventilation was offset by decreased residence time of the aerosol. We conclude that the Ultravent nebulizer can be used to nebulize drugs in a standard neonatal circuit, although the dose delivered is small. Tidal volume and aerosol residence time are important determinants of aerosol delivery.
吸入疗法对通气新生儿的益处已得到认可,但雾化器 - 呼吸机回路中药物输送的可靠性尚不确定。我们对改变变量的影响进行了量化。选用23只刚处死的兔子(体重1.15 - 1.9千克),通过气管切开术使用压力限制、时间切换的呼吸机(Neovent)进行通气。将来自Ultravent雾化器(Mallinkrodt)的高锝[99Tcm]放射性气雾剂注入呼吸机近端管道。在“标准设置”下进行两次3分钟的雾化,之后在改变压力、频率、气体流量、吸呼比或雾化器在回路中的位置的情况下再进行两次雾化。每次雾化后进行3分钟的γ相机成像,并测量切除的肺和气管中的总沉积放射性。图像显示外周气雾剂沉积良好。在标准设置下,肺沉积平均为释放气雾剂的2.8%。通过降低潮气量(呼吸机压力)和气雾剂停留时间(吸呼比),这一比例显著降低。气体流量降低会使沉积略有减少,可能是由于颗粒尺寸增大以及潮气量略有减少。沉积不会随频率增加而改变;分钟通气量增加被气雾剂停留时间减少所抵消。我们得出结论,尽管输送的剂量较小,但Ultravent雾化器可用于在标准新生儿回路中雾化药物。潮气量和气雾剂停留时间是气雾剂输送的重要决定因素。