Xu Xia, Roman John M, Veenstra Timothy D, Van Anda Jennifer, Ziegler Regina G, Issaq Haleem J
Laboratory of Proteomics and Analytical Technologies, SAIC-Frederick, Inc., National Cancer Institute at Frederick, Frederick, Maryland 21702, USA.
Anal Chem. 2006 Mar 1;78(5):1553-8. doi: 10.1021/ac051425c.
Packed column supercritical fluid chromatography with tandem mass spectrometry was used for the separation of estrone, estradiol, estriol, 16-epiestriol, 17-epiestriol, 16-ketoestradiol, 16alpha-hydroxyestrone, 2-methoxyestrone, 4-methoxyestrone, 2-hydroxyestrone-3-methyl ether, 2-methoxyestradiol, 4-methoxyestradiol, 2-hydroxyestrone, 4-hydroxyestrone, and 2-hydroxyestradiol. A gradient of methanol in carbon dioxide (0-30% methanol in 15 min, 2% change/min) at a flow rate of 2 mL/min and cyanopropyl silica column connected in series with a diol column, both 2.1 mm i.d. x 150 mm long, packed with 5-mum spherical silica-based particles, resulted in the separation and quantification of all 15 estrogens in less than 10 min. The limit of detection (LOD) and limit of quantitation (LOQ) of this pSFC MS/MS method was determined to be 0.5 (S/N = 3), and 5 pg, respectively. Compared with RP-HPLC MS analysis of the same mixture in terms of speed of analysis and sensitivity, pSFC MS is much faster, 10 versus 70 min, with comparable LOD and LOQ.
采用填充柱超临界流体色谱-串联质谱法分离雌酮、雌二醇、雌三醇、16-表雌三醇、17-表雌三醇、16-酮雌二醇、16α-羟基雌酮、2-甲氧基雌酮、4-甲氧基雌酮、2-羟基雌酮-3-甲基醚、2-甲氧基雌二醇、4-甲氧基雌二醇、2-羟基雌酮、4-羟基雌酮和2-羟基雌二醇。以2 mL/min的流速在二氧化碳中进行甲醇梯度洗脱(15分钟内甲醇含量从0%升至30%,每分钟变化2%),并使用内径2.1 mm、长度150 mm、填充5μm球形硅胶基颗粒的氰丙基硅胶柱与二醇柱串联,可在不到10分钟内实现所有15种雌激素的分离和定量。该超临界流体色谱串联质谱法的检测限(LOD)和定量限(LOQ)分别确定为0.5(信噪比=3)和5 pg。与对相同混合物进行的反相高效液相色谱-质谱分析相比,就分析速度和灵敏度而言,超临界流体色谱-质谱要快得多,分别为10分钟和70分钟,且检测限和定量限相当。