Suppr超能文献

Feasibility of ultra-low-dose multi-detector-row CT-colonography: detection of artificial endoluminal lesions in an in-vitro-model with optimization of image quality using a noise reduction filter algorithm.

作者信息

Branschofsky M, Vogt C, Aurich V, Beck A, Mödder U, Cohnen M

机构信息

Institute of Diagnostic Radiology, University Hospital Düsseldorf, Duesseldorf, Germany.

出版信息

Eur J Med Res. 2006 Jan 31;11(1):13-9.

Abstract

PURPOSE

To assess the most favorable slice thickness in Multi-Detector-Row CT-colonography (MDCTC), and the feasibility of dose reduction in an in-vitro-setting as well as the possibility of optimization of image quality using a noise reduction filter algorithm. -

MATERIALS AND METHODS

18 artificial lesions with sizes from 1 to 8 mm were randomly positioned in two cleansed pig colons. At a "Somatom Plus 4 Volume Zoom", six scanning protocols using a slice collimation of 2.5, 1, and 1 mm with a reconstructed slice thickness of 3, 3, and 1.25 mm were performed with tube currents of 100, and 10 mAs, respectively. Using a non-commercial software, a non-linear Gaussian filter was used to minimize image noise. Image noise was assessed before and after application of the filtering process. Using a threshold of -750 HU, two blinded readers analyzed the virtual colonography in respect to lesion location, size, and shape. Artifacts were noted. An automated detection system was evaluated. -

RESULTS

Using 10 mAs, a ten-fold dose reduction was achieved. After application of the mathematical filter, image noise was reduced by 45-80% for 100 mAs, and by 50-70% for 10 mAs scans. Only with a slice thickness of 1.25 mm, all lesions could be detected. The definition of lesion size and shape was more accurate with higher mAs. Only minor noise artifacts were noted on low-dose images. The automated polyp detector marked not more than 60% of artificial lesions. -

CONCLUSION

MDCTC benefits from narrow slice collimation. In an in-vitro-model, a significant dose reduction is achievable with preservation of a high lesion detection rate. The noise reduction filter algorithm improved image quality substantially.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验