Taylor Stuart A, Halligan Steve, Bartram Clive I, Morgan Paul R, Talbot Ian C, Fry Nicola, Saunders Brian P, Khosraviani Kirosh, Atkin Wendy
Department of Intestinal Imaging, St Mark's Hospital, Northwick Park, Watford Road, Harrow, Middlesex HA1 3UJ, England.
Radiology. 2003 Oct;229(1):109-18. doi: 10.1148/radiol.2291020561.
To investigate the effects of orientation, collimation, pitch, and tube current setting on polyp detection at multi-detector row computed tomographic (CT) colonography and to determine the optimal combination of scanning parameters for screening.
A colectomy specimen containing 117 polyps of different sizes was insufflated and imaged with a multi-detector row CT scanner at various collimation (1.25 and 2.5 mm), pitch (3 and 6), and tube current (50, 100, and 150 mA) settings. Two-dimensional multiplanar reformatted images and three-dimensional endoluminal surface renderings from the 12 resultant data sets were examined by one observer for the presence and conspicuity of polyps. The results were analyzed with Poisson regression and logistic regression to determine the effects of scanning parameters and of specimen orientation on polyp detection.
The percentage of polyps that were detected significantly increased when collimation (P =.008) and table feed (P =.03) were decreased. Increased tube current resulted in improved detection only of polyps with a diameter of less than 5 mm. Polyps of less than 5 mm were optimally depicted with a collimation of 1.25 mm, a pitch of 3, and a tube current setting of 150 mA; polyps with a diameter greater than 5 mm were adequately depicted with 1.25-mm collimation and with either pitch setting and any of the three tube current settings. Small polyps in the transverse segment (positioned at a 90 degrees angle to the z axis of scanning) were significantly less visible than those in parallel or oblique orientations (P <.001). The effective radiation dose, calculated with a Monte Carlo simulation, was 1.4-10.0 mSv.
Detection of small polyps (<5 mm) with multi-detector row CT is highly dependent on collimation, pitch, and, to a lesser extent, tube current. Collimation of 1.25 mm, combined with pitch of 6 and tube current of 50 mA, provides for reliable detection of polyps 5 mm or larger while limiting the effective radiation dose. Polyps smaller than 5 mm, however, may be poorly depicted with use of these settings in the transverse colon.
研究多排螺旋计算机断层扫描(CT)结肠成像中扫描方向、准直、螺距和管电流设置对息肉检测的影响,并确定用于筛查的扫描参数的最佳组合。
将含有117个不同大小息肉的结肠切除标本充气后,使用多排螺旋CT扫描仪在不同准直(1.25和2.5毫米)、螺距(3和6)和管电流(50、100和150毫安)设置下进行成像。由一名观察者检查从12个所得数据集中生成的二维多平面重组图像和三维腔内表面重建图像,以确定息肉的存在和清晰度。使用泊松回归和逻辑回归分析结果,以确定扫描参数和标本方向对息肉检测的影响。
当准直(P = 0.008)和床速(P = 0.03)降低时,检测到的息肉百分比显著增加。管电流增加仅改善了直径小于5毫米息肉的检测。直径小于5毫米的息肉在准直为1.25毫米、螺距为3且管电流设置为150毫安时显示最佳;直径大于5毫米息肉在准直为1.25毫米、任意螺距设置和三种管电流设置中的任何一种时均可充分显示。横段中的小息肉(与扫描z轴呈90度角)比平行或倾斜方向的息肉明显更难观察到(P < 0.001)。通过蒙特卡罗模拟计算的有效辐射剂量为1.4 - 10.0毫希沃特。
多排螺旋CT对小息肉(<5毫米)的检测高度依赖于准直、螺距,在较小程度上依赖于管电流。1.25毫米准直、螺距为6和管电流为50毫安的组合,可在限制有效辐射剂量的同时,可靠检测5毫米或更大的息肉。然而,在横结肠中使用这些设置时,小于5毫米的息肉可能显示不佳。