Curr Opin Ophthalmol. 2006 Apr;17(2):115-9. doi: 10.1097/01.icu.0000193093.05927.a1.
Central corneal thickness has become an important biometric factor and is an essential part of the evaluation of glaucoma. Goldmann applanation tonometry is the most widely used method of measuring intraocular pressure, but it is well known that corneal parameters affect the accuracy of this instrument. Intraocular pressure reduction is currently the only treatment available for decreasing the risk of glaucoma progression, so it is important to have an accurate reading of the true intraocular pressure.
New tonometers such as the dynamic contour tonometer and the Proview tonometer do not appear to be influenced by central corneal thickness. The ocular response analyzer provides a new measure of corneal biomechanics called corneal hysteresis. Rebound tonometry has been used in animal models of glaucoma, but has recently been compared to Goldmann applanation tonometry in humans. All of these technologies are presented in this review and the influence of corneal thickness on their accuracy is examined.
Confronted with the expanding evidence that central corneal thickness is an important ocular parameter that should be measured in clinical practice, eye-care professionals understandably wonder how to best obtain the measurements and what to do with the information. There is wide disagreement among investigators as to if there is an adequately validated 'correction algorithm'; without a validated algorithm, the argument goes, clinicians cannot use the data. These newer technologies may allow clinicians to be able to accurately determine the true intraocular pressure without having to rely on a correction algorithm.
中央角膜厚度已成为一项重要的生物测量因素,是青光眼评估的重要组成部分。Goldmann压平眼压计是测量眼压最广泛使用的方法,但众所周知,角膜参数会影响该仪器的准确性。降低眼压是目前降低青光眼进展风险的唯一可用治疗方法,因此准确读取真实眼压很重要。
动态轮廓眼压计和Proview眼压计等新型眼压计似乎不受中央角膜厚度的影响。眼反应分析仪提供了一种称为角膜滞后的角膜生物力学新测量方法。回弹眼压计已用于青光眼动物模型,但最近已在人体中与Goldmann压平眼压计进行了比较。本综述介绍了所有这些技术,并研究了角膜厚度对其准确性的影响。
面对越来越多的证据表明中央角膜厚度是临床实践中应测量的重要眼部参数,眼科护理专业人员理所当然地想知道如何最好地进行测量以及如何处理这些信息。关于是否存在经过充分验证的“校正算法”,研究人员之间存在广泛分歧;有人认为,没有经过验证的算法,临床医生就无法使用这些数据。这些新技术可能使临床医生能够准确确定真实眼压,而不必依赖校正算法。