Effect of central corneal thickness and corneal hysteresis on tonometry as measured by dynamic contour tonometry, ocular response analyzer, and Goldmann tonometry in glaucomatous eyes.

作者信息

Hager Annette, Loge Kristina, Schroeder Bernd, Füllhas Mark-Oliver, Wiegand Wolfgang

机构信息

Department of Ophthalmology, Asklepios Klinik Nord-Heidberg, Tangstedter Landstr. 400, 22419 Hamburg, Germany.

出版信息

J Glaucoma. 2008 Aug;17(5):361-5. doi: 10.1097/IJG.0b013e31815c3ad3.

Abstract

PURPOSE

The dynamic contour tonometer (DCT; Pascal tonometer) and the Ocular Response Analyzer (ORA) are novel tonometers designed to measure intraocular pressure (IOP) independent of corneal properties and central corneal thickness (CCT), respectively. We wanted to compare the corneal compensated IOP (IOPcc) as measured by ORA with IOP values measured by DCT and Goldmann applanation tonometry (GAT) with respect to IOP readings and the influence of corneal hysteresis (CH) and CCT in glaucoma patients.

PATIENTS AND METHODS

In a study of 94 glaucomatous eyes, IOP measurements by ORA, DCT, and GAT were compared, and the effects of CCT and CH were analyzed. All measurements were taken by 1 of the authors only.

RESULTS

The Mean CCT was 550+/-44 mum and mean CH 10.24+/-3.3 mm Hg. The mean value for IOPcc was 17.94+/-5.9 mm Hg, 15.14+/-3.7 mm Hg for DCT, and 14.3+/-4.3 mm Hg for GAT. The mean difference was 3.65+/-3.85 mm Hg between IOPcc and GAT and 2.80+/-4.9 mm Hg between IOPcc and DCT (P<0.001) and -0.85+/-3.3 mm Hg between GAT and DCT (P=0.015). None of the methods was related to CCT. In contrast, CH was highly statistically significant related to IOPcc (P<0.0001), whereas GAT and DCT showed only weak relation to CH (P=0.05).

CONCLUSIONS

IOP measurements with the ORA are significantly higher than DCT or GAT with no apparent linear correction factor. To date, we cannot estimate the impact of differences in CH in relation to IOP and its measurement.

摘要

目的

动态轮廓眼压计(DCT;帕斯卡眼压计)和眼反应分析仪(ORA)是新型眼压计,分别旨在测量独立于角膜特性和中央角膜厚度(CCT)的眼内压(IOP)。我们希望比较通过ORA测量的角膜补偿眼压(IOPcc)与通过DCT和Goldmann压平眼压计(GAT)测量的IOP值,涉及IOP读数以及角膜滞后(CH)和CCT对青光眼患者的影响。

患者与方法

在一项对94只青光眼患眼的研究中,比较了通过ORA、DCT和GAT进行的IOP测量,并分析了CCT和CH的影响。所有测量均仅由一位作者进行。

结果

平均CCT为550±44μm,平均CH为10.24±3.3mmHg。IOPcc的平均值为17.94±5.9mmHg,DCT为15.14±3.7mmHg,GAT为14.3±4.3mmHg。IOPcc与GAT之间的平均差值为3.65±3.85mmHg,IOPcc与DCT之间为2.80±4.9mmHg(P<0.001),GAT与DCT之间为-0.85±3.3mmHg(P=0.015)。没有一种方法与CCT相关。相比之下,CH与IOPcc具有高度统计学显著相关性(P<0.0001),而GAT和DCT与CH仅显示出微弱相关性(P=0.05)。

结论

使用ORA测量的IOP显著高于DCT或GAT,且没有明显的线性校正因子。迄今为止,我们无法估计CH差异对IOP及其测量的影响。