Dias Mauro S, Takeda Mauro N, Koskinas Marina F
Instituto de Pesquisas Energéticas e Nucleares: IPEN/CNEN-SP, Av. Prof. Lineu Prestes 2242, 05508-000, São Paulo, SP, Brazil.
Appl Radiat Isot. 2006 Oct-Nov;64(10-11):1186-92. doi: 10.1016/j.apradiso.2006.02.019. Epub 2006 Mar 23.
A technique for simulating all detection processes in a 4pi(beta,e,X)-gamma coincidence system by means of the Monte Carlo technique is described. This procedure yields a more realistic behaviour of the extrapolation curve as compared to the usual polynomial fit. The present paper describes its application to the standardisation of a typical pure beta emitter, namely 35S, by the efficiency tracing technique, and an EC-gamma radionuclide, namely 133Ba. The calculated extrapolations were compared to experimental values obtained at the IPEN.
描述了一种通过蒙特卡罗技术模拟4π(β,e,X)-γ符合系统中所有探测过程的技术。与通常的多项式拟合相比,该程序产生的外推曲线行为更符合实际情况。本文描述了其在通过效率追踪技术对典型的纯β发射体即35S以及电子俘获γ放射性核素即133Ba进行标准化方面的应用。将计算得到的外推值与在巴西核能与能源研究所(IPEN)获得的实验值进行了比较。