Walker Clare L, Whitby Liam, Granger Viv, Storie Ian, Reilly John T, Barnett David
UK NEQAS for Leucocyte Immunophenotyping, Department of Haematology, Royal Hallamshire Hospital, Sheffield.
Cytometry B Clin Cytom. 2006 May;70(3):154-62. doi: 10.1002/cyto.b.20103.
We have previously reported a flow rate calibration method for the determination of absolute CD4(+) T-lymphocyte counts that removes the need for the addition of latex beads to each sample. However, a limitation with this approach is that a calibration factor (CF) needs to be applied to adjust for differences in viscosity between latex bead suspensions and biological specimens. We have also demonstrated the value of using stabilized whole blood samples in external quality assessment (EQA) studies; such samples have a stable absolute lymphocyte count for over 1 year, at 4 degrees C. It was successfully demonstrated that this material can be used as a flow rate biocalibration (FRB) material for use as a flow cytometric control to provide a sample with a known CD4(+) T-lymphocyte count. Such material has advantages over latex bead technology as it can act as a full process control as well as having the same matrix and viscosity characteristics as the test material, thus removing the need for a CF.
In this study, we have analyzed 268 consecutive normal, abnormal, and HIV(+) samples using FRB, incorporating the PanLeucoGating approach and compared this to the MultiSet method, defined as the predicate.
Percentage similarity statistics revealed the following: 0-3,000 CD4(+) cells/mul mean percentage difference (MPD; bias) 1.2%, 95% CI of 5.6-8%; 0-200 CD4(+) cells/microl MPD of 1.25%, 95% CI of 11.63-14.13%; 201-500 CD4(+) cells/microl MPD of 1%, 95% CI of 4.6-6.6%.
This study demonstrates that stabilized whole blood can be used for FRB. It has the advantage of being a full process control, in addition to costing less than latex beads with highly comparable results. As bench top flow cytometers are extremely stable, this is a low cost and robust alternative to bead based methods for generating absolute CD4 counts.
我们之前报道了一种用于测定绝对CD4(+) T淋巴细胞计数的流速校准方法,该方法无需在每个样本中添加乳胶珠。然而,这种方法的一个局限性是需要应用校准因子(CF)来调整乳胶珠悬浮液和生物样本之间的粘度差异。我们还证明了在外部质量评估(EQA)研究中使用稳定化全血样本的价值;此类样本在4℃下绝对淋巴细胞计数可稳定超过1年。已成功证明这种材料可用作流速生物校准(FRB)材料,用作流式细胞术对照,以提供具有已知CD4(+) T淋巴细胞计数的样本。这种材料比乳胶珠技术具有优势,因为它可以作为全流程对照,并且具有与测试材料相同的基质和粘度特性,因此无需CF。
在本研究中,我们使用FRB分析了268个连续的正常、异常和HIV(+)样本,采用全白细胞设门方法,并将其与定义为判定标准的MultiSet方法进行比较。
百分比相似性统计显示如下:0 - 3000个CD4(+)细胞/μl,平均百分比差异(MPD;偏差)为1.2%,95%置信区间为5.6 - 8%;0 - 200个CD4(+)细胞/μl,MPD为1.25%,95%置信区间为11.63 - 14.13%;201 - 500个CD4(+)细胞/μl,MPD为1%,95%置信区间为4.6 - 6.6%。
本研究表明稳定化全血可用于FRB。它具有作为全流程对照的优势,此外成本低于乳胶珠,结果具有高度可比性。由于台式流式细胞仪极其稳定,这是一种低成本且可靠的替代基于珠子的方法来生成绝对CD4计数的方法。