Palmer H E, Rieksts G A, Lynch T P
Pacific Northwest Laboratory, Richland, WA 99352.
Health Phys. 1991 Nov;61(5):595-600. doi: 10.1097/00004032-199111000-00001.
An array of four large-volume coaxial germanium detectors was installed in a shielded room for the in vivo measurement of gamma emitters in the body. All four detectors are in a single 50-L dewar of liquid nitrogen and are positioned underneath the supine subject with the axes of the detector elements parallel with the length of the body. This orientation gives approximately 15% more counting efficiency than having the axes of the detectors perpendicular to the body length. The detectors have resolutions that are less than 2 keV with an average of 1.86 keV for the 60Co 1.33-MeV gamma-ray peak. The efficiencies of the four detectors range from 59% to 77% of a 7.62- x 7.62-cm (3- x 3-in.) NaI (Tl) detector. The detectors are interfaced to a multichannel analyzer such that spectra from the detectors are stored both as a sum of all counts from the four detectors and as a spectrum from each detector. The sensitivity of the germanium array for measuring radioactivity in the body exceeds that of multiple large NaI (Tl) detectors for measuring single radionuclides in the body and is much superior for measuring internal depositions of radionuclide mixtures. Another major advantage of this system compared to NaI(Tl) systems is that the results can be calculated by computer analysis with minimal human interpretation. The sensitivity of the array configuration is described and applications of the array in other configurations, such as that used for measuring the very small quantities of 24Na induced in the body during air travel, are also discussed.
在一个屏蔽室内安装了一组四个大体积同轴锗探测器,用于体内伽马射线发射体的测量。所有四个探测器都置于一个50升的液氮杜瓦瓶中,并放置在仰卧的受试者下方,探测器元件的轴与身体长度平行。这种取向比探测器轴垂直于身体长度时的计数效率高出约15%。对于60Co的1.33兆电子伏特伽马射线峰,探测器的分辨率小于2千电子伏特,平均为1.86千电子伏特。四个探测器的效率范围为一个7.62×7.62厘米(3×3英寸)碘化钠(铊)探测器效率的59%至77%。探测器与多道分析器相连,使得来自探测器的能谱既作为四个探测器所有计数的总和进行存储,也作为每个探测器的能谱进行存储。锗探测器阵列用于测量体内放射性的灵敏度超过多个大型碘化钠(铊)探测器用于测量体内单个放射性核素的灵敏度,并且在测量放射性核素混合物的体内沉积方面具有更大优势。与碘化钠(铊)系统相比,该系统的另一个主要优点是结果可以通过计算机分析计算得出,人工解读最少。文中描述了阵列配置的灵敏度,并讨论了该阵列在其他配置中的应用,例如用于测量空中旅行期间体内诱导产生的极少量24Na的配置。