Kahook M Y, Noecker R J, Pantcheva M B, Schuman J S
UPMC Eye Center, Eye and Ear Institute, University of Pittsburgh School of Medicine, Pittsburgh, PA 15213, USA.
Br J Ophthalmol. 2006 Aug;90(8):1010-3. doi: 10.1136/bjo.2006.091272. Epub 2006 Apr 13.
Limited data are available to guide optimal positioning of glaucoma drainage devices (GDD) in relation to the limbus and optic nerve. The authors aim to provide guidelines for appropriate and safe GDD implantation.
The optimal positioning of five different GDD were evaluated using necropsy eyes of varying axial lengths. The dependent variable that was measured was the maximum distance that a GDD could be placed posterior to the limbus while remaining 2 mm away from the optic nerve.
The average maximum distance posterior to the limbus of the anterior plate edge ranged between 9.0-15.0 mm in the superotemporal quadrant for the GDD tested. The distances for superonasal, inferonasal, and inferotemporal quadrants ranged between 8.0-14.0 mm, 9.0-14.0 mm, and 11.0-17.0 mm, respectively. The Molteno device could be placed most posteriorly while remaining 2 mm away from the nerve. The Ahmed FP7 and S2 were the least amenable to posterior placement before encroaching on the 2 mm limit.
The maximum distance that a GDD can be placed posterior to the limbus, before encroachment around the optic nerve, varies between different devices and quadrants of placement. Taking a measurement of the exact distance of the plate from the limbus during GDD surgery is recommended.
关于青光眼引流装置(GDD)相对于角膜缘和视神经的最佳定位,可用数据有限。作者旨在提供合适且安全的GDD植入指南。
使用不同眼轴长度的尸检眼评估五种不同GDD的最佳定位。所测量的因变量是GDD在角膜缘后方可放置的最大距离,同时与视神经保持2毫米的距离。
对于所测试的GDD,在前颞象限中,前板边缘在角膜缘后方的平均最大距离在9.0 - 15.0毫米之间。鼻上、鼻下和颞下象限的距离分别在8.0 - 14.0毫米、9.0 - 14.0毫米和11.0 - 17.0毫米之间。莫尔滕诺装置可以放置在最靠后的位置,同时与神经保持2毫米的距离。在接近2毫米限制之前,艾哈迈德FP7和S2最不适合向后放置。
在接近视神经之前,GDD在角膜缘后方可放置的最大距离因不同装置和放置象限而异。建议在GDD手术期间测量板与角膜缘的确切距离。