Suppr超能文献

Pacemaker-driven tachycardia induced by electrocardiograph monitoring in the recovery room.

作者信息

Hu R, Cowie D A

机构信息

Department ofAnaesthesia, Austin Health, Heidelberg, Victoria, Australia.

出版信息

Anaesth Intensive Care. 2006 Apr;34(2):266-8. doi: 10.1177/0310057X0603400221.

Abstract

Monitoring devices are known to induce tachycardia in minute-ventilation rate-responsive pacemakers. This is because some monitoring devices measure the same parameter as do the pacemakers (change in thoracic impedance). Hence, the biological signal to the pacemaker is increased and is misconstrued as increased minute ventilation causing tachycardia which resolves when the monitoring device is removed. Whilst this could occur for all minute-volume rate responsive pacemakers, most reported interactions have been with the Telectronics META series. We present a case of an interaction between a Telectronics Tempo DR pacemaker (St. Jude Medical) and an Agilent Patient Care System (Philips). Failure to recognise the true nature and cause of such tachycardias may lead to mismanagement of the patient, including the inappropriate use of cardio-active medications.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验