Preliminary results on correlation between neural response imaging and 'most comfortable levels' in cochlear implantation.

作者信息

Akin I, Kuran G, Saka C, Vural M

机构信息

SB Ankara Diskapi Training Hospital, 1st ENT Clinic, Ankara, Turkey.

出版信息

J Laryngol Otol. 2006 Apr;120(4):261-5. doi: 10.1017/S0022215106000442.

Abstract

During cochlear implantation, precise placement of the cochlear electrodes against the cochlear nerve is one of the most important steps, necessary to enable the patient to be aware of the stimulus from the implanted device. Neural response imaging (NRI) is a new tool which measures the evoked compound action potential generated as a result of electrical stimulation of the cochlea by properly placed electrodes during surgery. The aim of this study was to examine the correlation between intra-operative NRI recordings and 'most comfortable levels' (M levels) measured during standard fitting with the SoundWave fitting software, in cochlear implantation patients. Seven adult subjects were included in the study. The average duration of profound bilateral deafness was seven years (range = 19 years) and the mean age at implantation was 24 years (range = 17-34 years). All subjects underwent implantation and reached the one-month fitting session; one patient reached the two-month fitting session. The intra-operative NRI threshold (tNRI) was observed to be much higher than the first fitting M levels. The tNRI was 203 per cent of the sequential M value and 246 per cent of the paired M value. All but one intra-operative tNRI values were greater than 150 clinical units (CU), and all first fitting M levels were on average below 100 CU. No obvious correlation was found between individual intra-operative tNRI and first fitting M levels. The M levels at one month were much higher than the first fitting M levels, with much more dispersion.

摘要

在人工耳蜗植入过程中,将耳蜗电极精确放置在耳蜗神经上是最重要的步骤之一,这是使患者能够感知植入设备刺激所必需的。神经反应成像(NRI)是一种新工具,它测量在手术过程中电极正确放置时对耳蜗进行电刺激所产生的诱发复合动作电位。本研究的目的是检验人工耳蜗植入患者术中NRI记录与使用SoundWave拟合软件进行标准调试时测量的“最舒适水平”(M水平)之间的相关性。该研究纳入了7名成年受试者。双侧重度耳聋的平均持续时间为7年(范围 = 19年),植入时的平均年龄为24岁(范围 = 17 - 34岁)。所有受试者均接受了植入手术并进入了为期1个月的调试阶段;1名患者进入了为期2个月的调试阶段。术中观察到NRI阈值(tNRI)远高于首次调试时的M水平。tNRI为连续M值的203%,配对M值的246%。除1个值外,所有术中tNRI值均大于150临床单位(CU),而所有首次调试时的M水平平均低于100 CU。未发现个体术中tNRI与首次调试时的M水平之间存在明显相关性。1个月时的M水平远高于首次调试时的M水平,且离散度更大。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验