Cohen Mitchell I, Jedeikin Roy, Pfeiffer Sharon, Pedersen Scott
Department of Pediatric Cardiology, Phoenix Children's Hospital, 1920 E. Cambridge Avenue, Suite 301, Phoenix, AZ 85006, USA.
Europace. 2006 Apr;8(4):293-6. doi: 10.1093/europace/euj047. Epub 2006 Mar 14.
This report describes a 20-year-old man with complex congenital heart disease and inappropriate epicardial implantable cardioverter defibrillator (ICD) shocks secondary to double counting of ventricular-paced spikes and QRS complexes from a separate bipolar epicardial dual-chamber pacemaker. Adjusting to a unipolar paced-ventricular mode resolved any double counting via pacemaker-ICD interaction.