Suppr超能文献

计算机断层扫描定位像上的固有测量偏差不可预测:使用体模进行计算机断层扫描骨盆测量。

Intrinsic measurement bias on computed tomography scout view is unpredictable: Computed tomography pelvimetry using a phantom.

作者信息

Anderson N G, Fenwick J L, Wells J E

机构信息

Department of Radiology, Christchurch Hospital, Christchurch, New Zealand.

出版信息

Australas Radiol. 2006 Apr;50(2):127-31. doi: 10.1111/j.1440-1673.2006.01555.x.

Abstract

Our aim was to determine the degree of bias in CT scanogram measurements. We obtained standard lateral and anteroposterior (AP) pelvimetry scanograms of a phantom pelvis after placing ball bearings or aluminium rods to mark bony landmarks. Computed tomography pelvimetry was carried out at the manufacturer-recommended table height on two commercial CT scanners and at 10-mm increments up to 50 mm above and below this height. The AP inlet, AP outlet, interspinous distance and transverse diameters were each measured three times for each scanogram. The true measurements were obtained directly from the disassembled phantom. Bias was defined as the difference between the CT measurement and the true measurement. Observer error was negligible. The transverse diameter was overestimated at high table positions and underestimated at low table positions on both scanners (+6 to -10 mm). After correcting for geometric distortion, up to 6 mm bias was still present. The point at which no bias occurred was different for each scanner and did not correspond to the manufacturers' recommended table height. The outlet was overestimated on both scanners by up to 5 mm. The true inlet measurement was overestimated by 1.2 mm. The interspinous distance was minimally underestimated on both scanners. The measurements on CT scanogram were underestimated or overestimated in an inconsistent and unpredictable fashion, varying from one type of measurement to another and from CT scanner to CT scanner. This has implications for the accuracy and clinical utility of measurements obtained from a CT scanogram.

摘要

我们的目的是确定CT扫描测量中的偏差程度。在放置滚珠轴承或铝棒以标记骨性标志后,我们获得了一个模拟骨盆的标准侧位和前后位(AP)骨盆测量扫描图。在两台商用CT扫描仪上,于制造商推荐的检查床高度进行计算机断层扫描骨盆测量,并在该高度上下以10毫米的增量进行测量,最高可达50毫米。对每张扫描图的AP入口、AP出口、棘间距离和横径分别进行三次测量。真实测量值直接从拆解后的模拟模型中获得。偏差定义为CT测量值与真实测量值之间的差值。观察者误差可忽略不计。在两台扫描仪上,高检查床位置时横径被高估,低检查床位置时横径被低估(+6至-10毫米)。校正几何失真后,仍存在高达6毫米的偏差。无偏差出现的点因扫描仪而异,且与制造商推荐的检查床高度不一致。两台扫描仪上出口均被高估达5毫米。真实入口测量值被高估1.2毫米。两台扫描仪上棘间距离均被轻微低估。CT扫描图上的测量值以不一致且不可预测的方式被低估或高估,因测量类型和CT扫描仪的不同而有所变化。这对从CT扫描图获得的测量值的准确性和临床实用性具有影响。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验