Whinnett Zachary I, Davies Justin E R, Willson Keith, Chow Anthony W, Foale Rodney A, Davies D Wyn, Hughes Alun D, Francis Darrel P, Mayet Jamil
International Centre for Circulatory Health, National Heart and Lung Institute, St Mary's Hospital and Imperial College, 5, 9-61 North Wharf Road, London W2 1LA, UK.
Europace. 2006 May;8(5):358-66. doi: 10.1093/europace/eul017.
In this study, we apply non-invasive blood pressure (BP) monitoring, by continuous finger photoplethysmography (Finometer), to detect directly haemodynamic responses during adjustment of the atrioventricular (AV) delay of cardiac resynchronization therapy (CRT), at different heart rates.
Twelve patients were studied with six re-attending for reproducibility assessment. At each AV delay, systolic BP relative to a reference AV delay of 120 ms (SBPrel) was calculated. We found that at higher heart rates, altering the AV delay had a more pronounced effect on BP (average range of SBPrel=17.4 mmHg) compared with resting rates (average range of SBPrel=6.5 mmHg), P<0.0001. Secondly, peak AV delay differed between patients (minimum 120 ms, maximum 200 ms). Thirdly, small changes in AV delay had significant BP effects: programming AV delay 40 ms below the peak AV delay reduced SBPrel by 4.9 mmHg (P<0.003); having it 40 ms above the peak decreased SBPrel by 4.4 mmHg (P<0.0005). Finally, the peak AV delay is highly reproducible both on the same day and at 3 months (Bland-Altman difference: 3+/-8 ms).
Continuous non-invasive arterial pressure monitoring demonstrates that even small changes in AV delay from its haemodynamic peak value have a significant effect on BP. This peak varies between individuals, is highly reproducible, and is more pronounced at higher heart rates than resting rates.
在本研究中,我们应用通过连续手指光电容积脉搏波描记法(Finometer)进行的无创血压监测,直接检测在不同心率下心脏再同步治疗(CRT)的房室(AV)延迟调整期间的血流动力学反应。
对12例患者进行了研究,其中6例再次参与以进行重复性评估。在每个AV延迟时,计算相对于120毫秒的参考AV延迟的收缩压(SBPrel)。我们发现,与静息心率相比,在较高心率下改变AV延迟对血压的影响更为显著(SBPrel的平均范围=17.4 mmHg)(静息心率下SBPrel的平均范围=6.5 mmHg),P<0.0001。其次,患者之间的峰值AV延迟有所不同(最小值120毫秒,最大值200毫秒)。第三,AV延迟的微小变化对血压有显著影响:将AV延迟设置在峰值AV延迟以下40毫秒会使SBPrel降低4.9 mmHg(P<0.003);将其设置在峰值以上40毫秒会使SBPrel降低4.4 mmHg(P<0.0005)。最后,峰值AV延迟在同一天和3个月时具有高度可重复性(布兰德-奥特曼差异:3±8毫秒)。
连续无创动脉压监测表明,即使AV延迟与其血流动力学峰值之间的微小变化也会对血压产生显著影响。这个峰值在个体之间有所不同,具有高度可重复性,并且在较高心率下比静息心率下更为显著。