Brunner Patricia, Merwa Robert, Missner Andreas, Rosell Javier, Hollaus Karl, Scharfetter Hermann
Institute of Medical Engineering, Graz University of Technology, Graz, Austria.
Physiol Meas. 2006 May;27(5):S237-48. doi: 10.1088/0967-3334/27/5/S20. Epub 2006 Apr 24.
Magnetic induction tomography (MIT) of biological tissue is used for the reconstruction of the complex conductivity distribution kappa inside the object under investigation. It is based on the perturbation of an alternating magnetic field caused by the object and can be used in all applications of electrical impedance tomography (EIT) such as functional lung monitoring and assessment of tissue fluids. In contrast to EIT, MIT does not require electrodes and magnetic fields can also penetrate non-conducting barriers such as the skull. As in EIT, the reconstruction of absolute conductivity values is very difficult because of the method's sensitivity to numerical errors and noise. To overcome this problem, image reconstruction in EIT is often done differentially. Analogously, this concept has been adopted for MIT. Two different methods for differential imaging are applicable. The first one is state-differential, for example when the conductivity change between inspiration and expiration in the lung regions is being detected. The second one is frequency-differential, which is of high interest in motionless organs like the brain, where a state-differential method cannot be applied. An equation for frequency-differential MIT was derived taking into consideration the frequency dependence of the sensitivity matrix. This formula is valid if we can assume that only small conductivity changes occur. In this way, the non-linear inverse problem of MIT can be approximated by a linear one (depending only on the frequency), similar to in EIT. Keeping this limitation in mind, the conductivity changes between one or more reference frequencies and several measurement frequencies were reconstructed, yielding normalized conductivity spectra. Due to the differential character of the method, these spectra do not provide absolute conductivities but preserve the shape of the spectrum. The validity of the method was tested with artificial data generated with a spherical perturbation within a conducting cylinder as well as for real measurement data. The measurement data were obtained from a potato immersed in saline. The resulting spectra were compared with reference measurements and the preservation of the shape of the spectra was analyzed.
生物组织的磁感应断层扫描(MIT)用于重建被研究对象内部的复电导率分布κ。它基于对象对交变磁场的扰动,可用于电阻抗断层扫描(EIT)的所有应用,如功能性肺监测和组织液评估。与EIT不同,MIT不需要电极,并且磁场还可以穿透非导电屏障,如头骨。与EIT一样,由于该方法对数值误差和噪声敏感,重建绝对电导率值非常困难。为了克服这个问题,EIT中的图像重建通常采用差分法。类似地,这个概念也被应用于MIT。有两种不同的差分成像方法适用。第一种是状态差分,例如检测肺部区域吸气和呼气之间的电导率变化时。第二种是频率差分,在像大脑这样的静止器官中很有意义,因为在那里不能应用状态差分法。考虑到灵敏度矩阵的频率依赖性,推导了频率差分MIT的一个方程。如果我们可以假设只发生小的电导率变化,这个公式就是有效的。通过这种方式,MIT的非线性逆问题可以近似为一个线性问题(仅取决于频率),类似于EIT。牢记这个限制,重建了一个或多个参考频率与几个测量频率之间的电导率变化,得到归一化电导率谱。由于该方法的差分特性,这些谱不提供绝对电导率,但保留了谱的形状。用导电圆柱体内的球形扰动生成的人工数据以及实际测量数据对该方法的有效性进行了测试。测量数据是从浸泡在盐水中的土豆获得的。将得到的谱与参考测量进行比较,并分析谱形状的保留情况。