Ernau M C
Department of Biological Sciences and Neurobiology Research Center, State University of New York at Albany, Albany, New York 12222.
Plant Physiol. 1974 May;53(5):772-4. doi: 10.1104/pp.53.5.772.
Microelectrodes with a 1- to 2-mum tip diameter have been made which are capable of withstanding plant cell hydrostatic pressure on impalement. Filling the electrodes with 1% agar or 5% gelatin in 2 m KCl prevents cytoplasmic contents from moving into the electrode tip on impalement and therefore prevents the irreversible increase in resistance which often occurs. The agar and gelatin electrodes were tested in two fresh water algae, Nitella translucens and Mougeotia sp., and the potentials recorded were found comparable to those recorded with standard 2 m KCl electrodes.
已制作出尖端直径为1至2微米的微电极,这种电极在刺入时能够承受植物细胞的静水压力。在电极中填充2m KCl中的1%琼脂或5%明胶,可防止细胞质内容物在刺入时进入电极尖端,从而防止经常出现的电阻不可逆增加。琼脂电极和明胶电极在两种淡水藻类——透明丽藻和穆氏藻中进行了测试,发现记录到的电位与用标准2m KCl电极记录的电位相当。