ARC Research Group on Photosynthesis, Department of Botany, University of Sheffield, Sheffield, S10 2TN, United Kingdom.
Plant Physiol. 1983 Nov;73(3):534-41. doi: 10.1104/pp.73.3.534.
An apparatus is described which permits the simultaneous measurement of O(2) evolution and chlorophyll a fluorescence from illuminated discs or pieces of green leaves. O(2) is measured in the gas phase in a temperature-controlled chamber of approximately 5-milliliter capacity. Calibration is effected by injection of air through vents. Response time is approximately 1.5 seconds for O(2), and full scale deflection, in normal operating mode, is approximately 10 micromoles O(2). The apparatus may also be used to monitor fluorescence alone, in an open mode, in which gas is passed continuously through the chamber.
描述了一种装置,该装置允许同时测量受光的圆盘或绿色叶片的 O(2) 释放和叶绿素 a 荧光。O(2) 在一个约 5 毫升容量的温度控制室内的气相中进行测量。通过通风口注入空气来进行校准。O(2) 的响应时间约为 1.5 秒,在正常工作模式下,满刻度偏转为约 10 微摩尔 O(2)。该装置也可用于单独监测荧光,在开放模式下,气体连续通过腔室。