喉镜灯光的可靠性。

The reliability of laryngoscope lights.

作者信息

Howes B W

机构信息

Southmead Hospital, Westbury-on-Trym, Bristol BS10 5NB, UK.

出版信息

Anaesthesia. 2006 May;61(5):488-91. doi: 10.1111/j.1365-2044.2006.04611.x.

Abstract

The aim of this study is to compare the brightness of laryngoscope light between instruments, and the variability within the same instrument. A secondary aim is to compare the light from a reusable laryngoscope with a single use design. Testing equipment was constructed which provided a qualitative measure of light output from laryngoscopes. Anaesthetists assessed the minimum acceptable brightness for intubation using the intubating mannikin and a variable brightness laryngoscope. Laryngoscopes in anaesthetic rooms around the hospital were then tested repeatedly, and the results were compared to the anaesthetists' opinions. There was considerable intersubject and intrasubject variation in the reusable design, often failing to satisfy the minimum brightness. The single use blades were found to be brighter and more reliable than reusable blades.

摘要

本研究的目的是比较不同喉镜器械的灯光亮度,以及同一器械内的亮度差异。次要目的是比较可重复使用喉镜与一次性使用设计喉镜的灯光。构建了测试设备,以对喉镜的光输出进行定性测量。麻醉医生使用插管人体模型和可变亮度喉镜评估插管所需的最低可接受亮度。然后对医院麻醉室中的喉镜进行反复测试,并将结果与麻醉医生的意见进行比较。在可重复使用的设计中,受试者间和受试者内存在相当大的差异,常常无法满足最低亮度要求。发现一次性使用的刀片比可重复使用的刀片更亮且更可靠。

相似文献

1
The reliability of laryngoscope lights.
Anaesthesia. 2006 May;61(5):488-91. doi: 10.1111/j.1365-2044.2006.04611.x.
2
Minimum and optimum light output of Macintosh size 3 laryngoscopy blades: a manikin study.
Anaesthesia. 2007 Feb;62(2):163-8. doi: 10.1111/j.1365-2044.2006.04912.x.
3
Changes in the light intensity of the fiberoptic laryngoscope blade by steam sterilization.
Anesth Analg. 2007 Apr;104(4):908-10. doi: 10.1213/01.ane.0000255154.77505.c8.
4
Minimum and optimum light requirements for laryngoscopy in paediatric anaesthesia: a manikin study.
Anaesthesia. 2008 Jan;63(1):65-70. doi: 10.1111/j.1365-2044.2007.05259.x.
5
[Single-use laryngoscope blade assessment].
Ann Fr Anesth Reanim. 2004 Jul;23(7):694-9. doi: 10.1016/j.annfar.2004.03.017.
6
A comparison of plastic single-use with metallic reusable laryngoscope blades for out-of-hospital tracheal intubation.
Ann Emerg Med. 2007 Sep;50(3):258-63. doi: 10.1016/j.annemergmed.2007.04.022. Epub 2007 Jun 20.
9
Flexibility and light emission of disposable paediatric Miller 1 laryngoscope blades*.
Anaesthesia. 2006 Aug;61(8):792-9. doi: 10.1111/j.1365-2044.2006.04721.x.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验