一种用于调强放射治疗质量保证的实用蒙特卡罗监测单位验证工具。

A practical Monte Carlo MU verification tool for IMRT quality assurance.

作者信息

Fan J, Li J, Chen L, Stathakis S, Luo W, Du Plessis F, Xiong W, Yang J, Ma C-M

机构信息

Department of Radiation Oncology, Fox Chase Cancer Center, Philadelphia, PA 19111, USA.

出版信息

Phys Med Biol. 2006 May 21;51(10):2503-15. doi: 10.1088/0031-9155/51/10/010. Epub 2006 Apr 26.

Abstract

Quality assurance (QA) for intensity-modulated radiation therapy (IMRT) treatment planning and beam delivery, using ionization chamber measurements and film dosimetry in a phantom, is time consuming. The Monte Carlo method is the most accurate method for radiotherapy dose calculation. However, a major drawback of Monte Carlo dose calculation as currently implemented is its slow speed. The goal of this work is to bring the efficiency of Monte Carlo into a practical range by developing a fast Monte Carlo monitor unit (MU) verification tool for IMRT. A special estimator for dose at a point called the point detector has been used in this research. The point detector uses the next event estimation (NEE) method to calculate the photon energy fluence at a point of interest and then converts it to collision kerma by the mass energy absorption coefficient assuming the presence of transient charged particle equilibrium. The MU verification tool has been validated by comparing the calculation results with measurements. It can be used for both patient dose verification and phantom QA calculation. The dynamic leaf-sequence log file is used to rebuild the actual MLC leaf sequence in order to predict the dose actually received by the patient. Dose calculations for 20 patient plans have been performed using the point detector method. Results were compared with direct Monte Carlo simulations using EGS4/MCSIM, which is a well-benchmarked Monte Carlo code. The results between the point detector and MCSIM agreed to within 2%. A factor of 20 speedup can be achieved with the point detector method compared with direct Monte Carlo simulations.

摘要

使用电离室测量和模体中的胶片剂量测定法对调强放射治疗(IMRT)治疗计划和射束投送进行质量保证(QA)非常耗时。蒙特卡罗方法是放射治疗剂量计算中最准确的方法。然而,目前实施的蒙特卡罗剂量计算的一个主要缺点是其速度较慢。这项工作的目标是通过开发一种用于IMRT的快速蒙特卡罗监测单位(MU)验证工具,将蒙特卡罗方法的效率提高到实际可用的范围。本研究中使用了一种称为点探测器的特殊点剂量估计器。点探测器使用下一个事件估计(NEE)方法来计算感兴趣点处的光子能量注量,然后通过质量能量吸收系数将其转换为碰撞比释动能,前提是存在瞬态带电粒子平衡。通过将计算结果与测量结果进行比较,对MU验证工具进行了验证。它可用于患者剂量验证和模体QA计算。动态叶片序列日志文件用于重建实际的多叶准直器(MLC)叶片序列,以便预测患者实际接受的剂量。使用点探测器方法对20个患者计划进行了剂量计算。结果与使用EGS4/MCSIM进行的直接蒙特卡罗模拟结果进行了比较,EGS4/MCSIM是一个经过充分验证的蒙特卡罗代码。点探测器和MCSIM之间的结果在2%以内相符。与直接蒙特卡罗模拟相比,点探测器方法可实现20倍的加速。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验